Operation
pH/Oxi 340i
22
Note
You can automatically print a calibration protocol after the calibration.
To do so, connect a printer to the interface according to section 4.7.2
before calibrating. After a valid calibration, the record is printed.
Sample printout:
Calibration evaluation
After the calibration, the measuring instrument automatically evaluates
the calibration. The asymmetry and slope are evaluated separately.
The worst evaluation appears on the display.
CALIBRATION PROTOCOL
02.03.01
14:19
Device No.: 12345678
Calibration pH
Cal time: 01.03.01 / 15:20
Cal interval:
7d
AutoCal TEC
Tauto
Buffer 1
2.00
Buffer 2
4.01
Buffer 3
7.00 *
Buffer 4
10.00
C1
184.1 mV
25.0°C
C2
3.0 mV
25.0°C
S1
-59.4 mV/pH
ASY1
-
4 mV
Probe:
+++
Display
Asymmetry
[mV]
Slope
[mV/pH]
-15 ... +15
-60.5 ... -58
-20 ... +20
-58 ... -57
-25 ... +25
-61 ... -60.5
or
-57 ... -56
Clean the electrode according to the
electrode operating manual
-30 ... +30
-62 ... -61
or
-56 ... -50
E3
Eliminate the error according to chap-
ter 6 W
HAT
TO
DO
IF
...
< -30 or
> 30
... -62 or
... -50