Oxi 197i
Operation
25
4.3.2
Calibrating via a comparison measurement
Proceed as follows to calibrate the instrument:
1
Connect the D. O. sensor to the measuring instrument.
2
Press the
<CAL>
key repeatedly until
O
2
FAC
appears.
3
Immerse the D. O. sensor in the calibrating solution.
4
Press the
<RUN/ENTER>
key.
The AutoRead measurement begins. If the measured value is
stable, the instrument displays the determined measured value
and the correction factor of the relative slope (initial value =
1.000).
5
Remove the D. O. sensor from the calibrating solution.
6
Perform a comparison measurement, e. g. Winkler titration,
with the calibrating solution (determine the nominal value).
Leave the measuring instrument switched on until the nominal
value is determined.
7
Set the displayed value on the measuring instrument to the
nominal value with
<
▲
> <
▼
>
. This changes the related correc-
tion factor of the relative slope.
REL1
S
AR
Year
Ident
No.
Day.Month
Tref25
Tref20
LoBat
Lin
nLF
mg/ l
ARng
Sal
pH
1
TEC
Auto
Store
8
Time
Baud
Sal
cm
1/
K
/
%
RCL
REL2
pH
/
mV
m
cm
S/
%
F
1 7 6
A C
°
C
TP
O
Oxi
Cal
TP
REL1
S
Year
Ident
No.
Day.Month
Tref25
Tref20
LoBat
Lin
nLF
Sal
pH
1
TEC
Auto
Store
Time
Baud
Sal
cm
1/
K
/
%
RCL
REL2
pH
/
mV
m
cm
S/
%
4
0 0 0
4 2
°
C
O
1
AR
ARng
Oxi
Cal
mg/ l
Содержание Oxi 197i
Страница 46: ...Operation Oxi 197i 46 ...
Страница 52: ...What to do if Oxi 197i 52 ...
Страница 56: ...Technical data Oxi 197i 56 ...
Страница 62: ...Index Oxi 197i 62 ...