Oxi 197i
Operation
23
4.3.1
Calibrating with the air calibration vessel
Proceed as follows to calibrate the instrument:
Note
The sponge in the air calibration vessel must be moist (not wet). Follow
the instructions in the OxiCal
®
-SL or DurOx 325 operating manual.
1
Connect the D. O. sensor to the measuring instrument.
2
Keep an air calibration vessel ready for use.
3
Put the D. O. sensor into the air calibration vessel.
4
Press the
<CAL>
key repeatedly until
O
2
CAL
appears.
5
Press the
<RUN/ENTER>
key. AutoRead is active,
AR
flashes.
6
As soon as a stable value is achieved,
AR
stops flashing. Cal-
ibration is finished; the sensor symbol indicates the relative
slope determined and the sensor evaluation (see page 22).