Operation
Oxi 170
104
ba24105e08
11/2010
4.9
Test mode
General information
Test mode can be used for the following purposes:
z
Set specific current values on the current outputs for test purposes
z
Switch the relays on and off for test purposes (RT and RT RS
option)
z
Check the RS 485 communication (RT RS option)
Test procedure
1
Keep
<ESC>
pressed and briefly
press
<UP>
at the same time.
Note: You can leave the test mode
at any time again with
<M>
or
<C>
.
2
Press
<ENTER>
.
The set password is displayed
("000" = password protection
deactivated).
3
Proceed to the next test item with
<ENTER>
.
Test of current output 1.
4
Press
<ENTER>
.
A fixed current value can be set on
current output 1 here:
– Set 0.1 mA with
<M>
– Set 20.0 mA with
<C>
– The value can be increased or
decreased as required
(adjustment range 0.0 to
20.5 mA) with
<UP><DOWN>
5
Proceed to the next test item with
<ENTER>
.
8
AR
REL1
0 0 0
REL2
C o d
8
AR
REL1
REL2
C o d
2 3
1
AR
REL1
REL2
C 1
e
r
t s t
C
AR
REL1
REL2
=1
C 1
e
r
Содержание Oxi 170
Страница 1: ...ba24105e08 11 2010 Oxi 170 Operating instructions Oxygen Monitor Oxi 170 O mg l 1 1 85 C REL1 5 6 ...
Страница 8: ...Overview Oxi 170 70 ba24105e08 11 2010 ...
Страница 18: ...Installation Oxi 170 80 ba24105e08 11 2010 ...
Страница 46: ...Maintenance and cleaning Oxi 170 108 ba24105e08 11 2010 ...
Страница 48: ...What to do if Oxi 170 110 ba24105e08 11 2010 ...
Страница 50: ...Accessories Oxi 170 112 ba24105e08 11 2010 ...
Страница 62: ...Index Oxi 170 124 ba24105e08 11 2010 ...
Страница 63: ......