Oxi 170
Operation
87
ba24105e08
11/2010
CR1
z
rEC
mg/l / %
Measured value range of the REC 1 analog
output (D.O. value).
After confirming the
reC
(mg/l or %) selection with
<ENTER>
you can
adapt the upper final value.
Depending on the measuring range, the
following settings are possible:
Measuring range 0.0 ... 60.0 mg/l:
(step size 5 ... mg/l)
Measuring range 0 ... 600 %:
(step size 100 %)
Measuring range 0.00 ... 20.00 mg/l:
(step size 1 ... mg/l)
Measuring range 0.0 ... 200.0 %:
(step size 10 %)
Note: If the range is exceeded, the recorders
output a constant value of 20.5 mA (
OFL
display). The instrument automatically
returns to the defined recorder range when
the value is within the range again.
Setting
Selection/Values
Explanation
(values in bold typeface = values on delivery)
Upper final value:
5 ...
60
mg/l
Lower final value:
0,0 mg/l
Upper final value:
100 ...
600
%
Lower final value:
0 %
Upper final value:
2 ...
20
mg/l
Lower final value:
0.00 mg/l
Upper final value:
100.0 ...
200.0
%
Lower final value:
0.0 %
Содержание Oxi 170
Страница 1: ...ba24105e08 11 2010 Oxi 170 Operating instructions Oxygen Monitor Oxi 170 O mg l 1 1 85 C REL1 5 6 ...
Страница 8: ...Overview Oxi 170 70 ba24105e08 11 2010 ...
Страница 18: ...Installation Oxi 170 80 ba24105e08 11 2010 ...
Страница 46: ...Maintenance and cleaning Oxi 170 108 ba24105e08 11 2010 ...
Страница 48: ...What to do if Oxi 170 110 ba24105e08 11 2010 ...
Страница 50: ...Accessories Oxi 170 112 ba24105e08 11 2010 ...
Страница 62: ...Index Oxi 170 124 ba24105e08 11 2010 ...
Страница 63: ......