What to do if...
7 - 1
ba75397d03
12/2004
7
What to do if...
No measured value
Measurement does not
function
Measurement provides
implausible measured
values
System cannot be
calibrated
Cause
Remedy
– Sensor not connected
– Connect the sensor
– Unknown
– Look in the log book
Cause
Remedy
– Electrode not connected
– Connect electrode
– Liquid has penetrated the
sensor
– Sensor defective, send it back
– Sensor not connected
– Connect the sensor
– Instrument setting incorrect
– Correct instrument setting
Cause
Remedy
– No calibration performed
– Calibrate
– Electrode not connected or
defective
– Check electrode and electrode
connection
– Electrode contaminated
– Clean electrode
– Liquid has penetrated the
sensor
– Sensor defective, send it back
– Instrument setting incorrect
– Correct the instrument setting
(
Measuring mode
mg/l or mV)
– Chloride compensation is
switched off
– Chloride compensation with
unsuitable value for chloride
content
– Switch on chloride compensa-
tion
– Determine and enter chloride
content
– Recalibrate (see also section
Cause
Remedy
– Slope of the electrode not
within tolerance (see section
4.1.10)
– Condition the electrode
– If the slope is still outside the
tolerance: Replace the elec-
trode
Содержание NitraLyt 700 IQ
Страница 1: ...ba75397e03 12 2004 NitraLyt 700 IQ Operating manual IQ SENSOR NET nitrate sensor...
Страница 10: ...Safety NitraLyt 700 IQ 2 4 ba75397d03 12 2004...
Страница 20: ...Commissioning NitraLyt 700 IQ 3 10 ba75397d03 12 2004...
Страница 56: ...Technical data NitraLyt 700 IQ 8 4 ba75397d03 12 2004...
Страница 60: ...Indexes NitraLyt 700 IQ 9 4 ba75397d03 12 2004...