Calibration and measuring
NitraLyt
®
700 IQ
4 - 20
ba75397d03
12/2004
4.1.10 Calibration result
Calibration evaluation
After calibrating the system automatically evaluates the calibration data
and current state of the sensor. The drift potential and slope are evalu-
ated separately. For a valid calibration, the values have to be within the
following ranges:
Add standard to sample
Confirm with
g
.
* Wait for a stable measured
value.
Confirm with
g
.
The measurement begins.
Cal.: DOUBLE STD. ADD. (5)
Reference voltage determined
after first standard addition
Confirm with
g
.
Add standard to sample
Confirm with
g
.
* Wait for a stable measured
value.
Confirm with
g
.
The measurement begins.
Cal.: DOUBLE STD. ADD. (5)
Reference voltage determined
after second standard addition
Confirm with
g
.
Calibration successful
Conc. (NO3-N)
x mg/l
Slope
y mV*
Drift voltage
z mV
End of DOUBLE STD ADD (5) cal.
The values for
Conc. (NO3-N)
,
Slope
and
Drift voltage
are dis-
played.
Calibration is complete. Confirm
with
g
.
The display returns to the mea-
sured value display.
Display
Explanation
Slope:
-50 ... -70 mV
Drift potential:
-100 ... +100 mV
Содержание NitraLyt 700 IQ
Страница 1: ...ba75397e03 12 2004 NitraLyt 700 IQ Operating manual IQ SENSOR NET nitrate sensor...
Страница 10: ...Safety NitraLyt 700 IQ 2 4 ba75397d03 12 2004...
Страница 20: ...Commissioning NitraLyt 700 IQ 3 10 ba75397d03 12 2004...
Страница 56: ...Technical data NitraLyt 700 IQ 8 4 ba75397d03 12 2004...
Страница 60: ...Indexes NitraLyt 700 IQ 9 4 ba75397d03 12 2004...