Multi 9420
Dissolved oxygen (D.O.)
ba75964e04
04/2015
75
7.3.6 Calibration data
Displaying the
calibration record
The calibration record of the last calibration is to be found under the menu item,
Calibration
/
Calibration record
. To open it in the measured value display, press
4.
Start the measurement with
<ENTER>
.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
5.
Wait for the end of the AutoRead measurement or accept the calibration
value with
<ENTER>
.
The factor that was set last is displayed.
6.
Using
<
> <
>
, set the correction factor to adjust the displayed con-
centration value to the nominal value (value of the comparison mea-
surement). Subsequently, accept the correction factor with
<ENTER>
.
The meter switches to the measured value display.
The status display
[Factor]
is active.
The calibration record is automatically transmitted to the interface
after calibrating.
03.04.2013 08:00
03.04.2013 08:00
Содержание Multi 9420
Страница 1: ...OPERATING MANUAL ba75964e04 04 2015 Multi 9420 DIGITAL METER FOR IDS SENSORS...
Страница 91: ...Multi 9420 Settings ba75964e04 04 2015 91...
Страница 113: ...Multi 9420 Maintenance cleaning disposal ba75964e04 04 2015 113...
Страница 131: ......