Multi 9420
Ion concentration
ba75964e04
04/2015
51
Continuing with two-
point calibration
9.
Immerse the electrode in standard solution 1.
10.
When calibrating without temperature sensor:
Measure the temperature of the standard solution using a thermom-
eter.
Use
<F4>
/[
⇅
] to select the setting of the temperature.
Use
<
> <
>
to set the temperature.
Use
<F4>
/[
⇅
] to select the setting of the concentration.
11.
Set the concentration of the standard solution with
<
> <
>
and
press
<ENTER>
.
The standard solution is measured.
The measured value is checked for stability (AutoRead).
12.
Wait for the end of the AutoRead measurement or accept the calibra-
tion value with
<ENTER>
.
The calibration display for the next standard solution appears.
01.08.2013 08:00
01.08.2013 08:00
Standard
13.
Thoroughly rinse the electrode with distilled water.
14.
Immerse the electrode in standard solution 2.