Technical data
MIQ/24V
5 - 2
ba64114e02
08/2006
Caution
If the original fuse needs to be replaced, it may only be replaced
by a fuse of the same type and manufacturer.
Instrument safety
Terminal connections
Caution
Only the line cross-sections specified under E
LECTRICAL
DATA
,
Seite 1 can be used for the 24 V AC/DC connection.
Number of MIQ/24V in
an IQ S
ENSOR
N
ET
sys-
tem
Maximum of 3, depending on the overall
power requirement of the system (see sys-
tem operating manual, section P
RINCIPLES
OF
AN
OPTIMUM
INSTALLATION
)
Instrument internal
fuses
5 x 20 T 1.6 A, with UL approval
Manufacturer:
Littlefuse
Type:
Series 218
Applicable norms
–
EN 61010-1
–
UL 3111-1
–
CAN/CSA C22.2 No. 1010.1
IQ S
ENSOR
N
ET
connec-
tions
3
Additional connectable SENSORNET termi-
nator (terminating resistor)
Terminal type
Screw-type terminal strip, accessible by
raising the lid
Terminal ranges
Solid wires:
0.2 ... 4.0 mm
2
AWG 24 ... 12
Flexible wires:
0.2 ... 2.5 mm
2
Cable feeds
4 cable glands M16 x 1.5 on the underside
of the module