LF 170
Operation
97
ba54104e09
11/2010
skipped.
z
From any point of the main level press
<ESC>
to return to the
starting point of the parameterization level. From here change to the
operating level with
<M>
or
<ESC>
or to the configuration level with
<UP><DOWN>
.
z
From any point of the main level press
<M>
to also go directly to the
operating level.
z
In view mode you can change to the operating level from any point
with
<M>
.
Setting table:
Setting
Selection/Values
Explanation
(values in bold typeface = values on delivery)
PR1
z
4
:
20
z
0
:
20
or
dI
:
=1
...
dI
:
20
Current range and attenuation of the REC 1
current output in
rEC
operating mode
(analog output).
Operating notes:
– The setting is carried out in the order:
current range -> attenuation
– Press 1x
<ENTER>
in between.
The following values can be set:
z
Current range:
–
4 to 20 mA
, or
– 0 to 20 mA
z
"dI" = attenuation dI/dt (in mA/s):
– 0.1
– 1
– 5
– 20
Attenuation is the change speed of the
current signal in the case of erratic
changes of the input signal.
PR2
(only with RT and
RT RS option)
z
4
:
20
z
0
:
20
or
dI
:
=1
...
dI
:
20
Current range and attenuation of the REC 2
current output in
rEC
operating mode
(analog output).
For setting possibilities, see
Pr1
(
ReC
operating mode).