Operation
LF 170
92
ba54104e09
11/2010
CR1
(continued)
The following settings are possible:
, Measuring range 0.000 ... 1.999 µS/cm:
(fixed range)
, Measuring range 0,00 ... 19,99 µS/cm:
(step size 1 µS/cm)
, Measuring range 0.0 ... 199.9 µS/cm:
(step size 10 µS/cm)
, Measuring range 0 ... 1999 µS/cm:
(step size 100 µS/cm)
, Measuring range 0.00 ... 19.99 mS/cm:
(step size 1 mS/cm)
(continues on the next page)
Setting
Selection/Values
Explanation
(values in bold typeface = values on delivery)
Upper final value:
0,000 µS/cm
Lower final value:
1.999 µS/cm
Upper final value:
2.00 ...
19.99
µS/cm
Lower final value:
0.00 ... 15.00 µS/cm
Minimum spacing:
2 µS/cm
Upper final value:
20.0 ...
199.9
µS/cm
Lower final value:
0.0 ... 150.0 µS/cm
Minimum spacing:
20 µS/cm
Upper final value:
500 ...
1999
µS/cm
Lower final value:
0 ... 1500 µS/cm
Minimum spacing:
500 µS/cm
Upper final value:
5.00 ...
19.99
mS/cm
Lower final value:
0.00 ... 14.99 mS/cm
Minimum spacing:
5 mS/cm