K 500 / K 800

Aging

ba15360e03

10/2014

7

Aging

Please note that every (combination) electrode undergoes a natural aging 
process.The response time increases and the slope decreases with the age 
of the (combination) electrode.The following factors shorten the lifetime con-
siderably:

Incorrect storage

Special measuring conditions (e.g. organic solutions, frequent measure-
ment with high concentrations of interfering ions) 

High temperatures

High changes in temperature 

The warranty does not cover damage caused by measuring conditions and 
mechanical damage.

Maintenance

Combination electrodes:
Refill any used up bridge electrolyte.

Install an exchange membrane cap.

NOTE

Never pull the connection cable of the electrode.
The cable might be damaged.

Storage

Between two

measurements

Condition the electrode in standard solution.

Overnight

K 500: 

Rinse the electrode with deionized water and store it in a dry place.

K 800: 

Rinse the combination electrode with deionized water and dry it. Then 

put on the storage vessel and store the electrode upright.

Longer than over-

night

K 500: 

Rinse the electrode with deionized water, then dab it dry with a clean 

paper towel and put on the protective cap. Store the electrode in a dry place.

K 800: 

Remove the bridge electrolyte, then rinse the combination electrode 

with deionized water and dab it dry with a clean paper towel. Store the combi-
nation electrode in the storage vessel in a dry place. For the following 
measurement, prepare the combination electrode as described in section 
C

OMMISSIONING

.

Store the reference electrode according to the instructions in its operating 
manual.

Содержание K 500 BNC

Страница 1: ...OPERATING MANUAL ba15360e03 10 2014 K 500 DIN K 500 BNC K 800 DIN K 800 BNC POTASSIUM ION SENSITIVE ELECTRODE...

Страница 2: ...f the present operating manual can be found on the Internet under www WTW com It is also available in other languages Copyright 2014 WTW GmbH Reprinting even as excerpts is only allowed with the expli...

Страница 3: ...3 View 1 Connection head with connection cable 2 Shaft 3 Exchange membrane cap 4 PVC membrane 5 O ring 6 Inner junctions 7 Exterior junctions 8 Storage vessel for K 800 9 Protective cap for K 500 1 2...

Страница 4: ...orage vessel 2 Fill ELY BR 503 K bridge electrolyte into the filling opening so that the inner junctions are covered with bridge electrolyte 3 Rinse the combination electrode with deionized water 4 Wi...

Страница 5: ...he meter The ion sensitive elec trode is sufficiently conditioned when the measured value no longer changes considerably under otherwise constant conditions temperature composition of the solution Min...

Страница 6: ...erning sample prep aration and measuring procedures WTW provides a large number of appli cation reports for various applications Response times The response time depends on the concentration range It...

Страница 7: ...ectrode The cable might be damaged Storage Between two measurements Condition the electrode in standard solution Overnight K 500 Rinse the electrode with deionized water and store it in a dry place K...

Страница 8: ...ination electrode K 800 K 800 AT 106672 Reference electrode for potassium electrode K 500 R 503 P R 503 D 106570 106571 Bridge electrolyte ELY BR 503 K 106577 ISA sample conditioning solution for K me...

Страница 9: ...rolyte Junctions encrusted Leave the bridge electrolyte to react on the junctions for some hours until the crusts have dissolved Cable broken Exchange the electrode Cause Remedy Conditioning time too...

Страница 10: ...ature range Membrane resis tance Length Diameter Cable length Plug DIN plug or BNC plug depending on design 0 04 39 000 mg l K 1 x 10 6 1 mol L K 2 2 12 see INTERFERENCES 0 40 C 50 M K 500 170 mm incl...

Отзывы: