inoLab pH/ION 735
Operation
57
8
Using
82
and
g
, enter the values for the volume of the
sample (
Sample volume
), the volume of the ISA/TISAB solu-
tion (
ISA/TISAB vol.
), the volume of the blank value solution
(
BV volume
) and the concentration of the blank value solution
(
BV concentration
).
9
Select
Continue
with
82
and start the measurement with
g
.
When the measurement is finished an info window appears.
10
Add the standard solution to the sample.
11
Start the measurement with
g
.
When the measurement is finished an entry window appears.
12
Using
82
and
g
, enter the values for the volume of the
standard solution (
Std. volume
) and concentration of the stan-
dard solution (
Std. concentration
).
13
Select
Continue
with
82
and confirm with
g
.
The measurement result is displayed.
14
If necessary, start measuring further samples with
g
.
Repeat steps 6 - 13 for all samples.
i
Add standard
i
U = 22.0 mV
Continue
Blank value addition
Std. volume
100.0 ml
Std. concentration
1.0 mg/l
Continue
Blank value addition
ISE
Blank value addition
0.321
mg/l
24.8
°C
20.09.04 10:37