Calibration points
Calibration can be performed using one, two or three buffer solutions in
any order (single-point-, two-point or three-point calibration). The meter
determines the following values and calculates the calibration line as
follows:
Note
You can display the slope in the units, mV/pH or % (see section 4.3.2).
AutoRead
In calibration, the AutoRead function is automatically activated.
The current AutoRead measurement can be terminated at any time
(accepting the current value).
Calibration record
When finishing a calibration, the new calibration values are displayed
as an informative message (
i
symbol) first. Then you can decide
whether you want to take over these values of the new calibration or
whether you want to continue measuring with the old calibration data.
After accepting the new calibration values the calibration record is dis-
played.
Display calibration data
and output to interface
You can have the data of the last calibration displayed (see
section 4.6.5). Subsequently, you can transmit the displayed calibra-
tion data to the interface, e. g. to a printer or PC, with the
p
key.
Note
The calibration record is automatically transmitted to the interface after
calibrating.
Determined val-
ues
Displayed calibration data
1-point
ASY
!
Asymmetry =
ASY
!
Slope = Nernst slope
(-59.16 mV/pH at 25 °C)
2-point
ASY
SLO
!
Asymmetry =
ASY
!
Slope =
SLO
3-point
ASY
SLO
!
Asymmetry =
ASY
!
Slope =
SLO
The calibration line is calculated
by linear regression.