Operation
107
4.3.2 AutoCal TEC
Use any two of the WTW technical buffer solutions for this
procedure in increasing or decreasing order.
Note
Steps 3 and 7 are not required if you use a temperature
probe.
1 Press the
m
key repeatedly until the pH/Redox
measurement appears on the display. pH or U
appears in the status line.
2 Press the
c
key repeatedly until the AutoCal TEC
function display appears.
3 If necessary, set the temperature of the buffer
solution using
u d
while keeping the
g
key
depressed.
4 Submerse the pH electrode in the first buffer solution.
5 Press the
g
key.
AR flashes on the display. The electrode voltage
(mV) appears on the display.
As soon as a stable value is recognized, Ct2
appears.