DIQ/S 181
Outputs
75
ba77083e01
02/2015
6.4
Setting table for relays
6.4.1
Functions and settings
To set a relay, its function has to be selected first. Then the relevant
setting table is displayed:
Relay function and
relevant setting tables
6.4.2
System monitoring
Function
The
System monitoring
enables the monitoring of system errors.
To set up the
System monitoring
function for a relay output, you have
to select the
System
option when linking the relay output (see
section 6.2.1).
It can be used to monitor the following system errors:
Settings
Function
Setting table
No function
The relay output is not used.
System monitoring
Sensor monitoring
Limit indicator
Frequency controller
Pulse-width output
Cleaning
Manual control
Setting
Selection
Explanation
Power failure
On
Off
The
Power failure
On
function
monitors the supply voltage. If
the voltage falls below a critical
value, the relay switches.
Collective error
On
Off
The
Collective error On
function simultaneously
monitors the proper function of
the sensor. (for details, see
below this table)
Action
Open
For all functions of the
System
monitoring
, the relay action is
set to
Open
.
Содержание DIQ/S 181
Страница 1: ...OPERATING MANUAL ba77083e01 02 2015 DIQ S 181 UNIVERSAL TRANSMITTER FOR ONE DIGITAL SENSOR...
Страница 2: ...Copyright 2016 Xylem Analytics Germany GmbH Printed in Germany...
Страница 6: ...List of contents DIQ S 181 6 ba77083e01 02 2015...
Страница 102: ...Index DIQ S 181 102 ba77083e01 02 2015...
Страница 103: ......