Outputs
DIQ/S 181
76
ba77083e01
02/2015
Collective error
message
In the case of a
Collective error
message, the relay opens if one of the
following malfunctions occurs:
The sensor, which is properly registered to the DIQ/S 181, does not
provide a valid main measured value
The sensor, which is properly registered to the DIQ/S 181, does not
provide a valid secondary measured value
In any case, the relay remains open for 10 seconds and only closes
when the malfunction is no longer present.
In the following cases, the relay does not open despite an invalid
measured value:
The sensor is being calibrated
The sensor is in the maintenance condition
The sensor is being cleaned with the aid of a valve module
(compressed air operated cleaning system).
6.4.3
Sensor monitoring
Function
The
Sensor monitoring
function enables the monitoring of sensor errors
and the maintenance condition.
In order to set up the
Sensor monitoring
function for a relay output, the
relay output must be linked with the sensor (see section 6.2.1).
Settings
Setting
Selection
Explanation
Errors
All
All sensor errors (special ones
and general ones) are
monitored and can prompt a
relay action.
Special
Special sensor errors are
monitored and can prompt a
relay action.
Off
Sensor errors are not
monitored.
Mainten. Condition
On
Off
Switching on and off the
maintenance condition is
monitored and can prompt a
relay action.
Action
Open
Close
Relay action
Содержание DIQ/S 181
Страница 1: ...OPERATING MANUAL ba77083e01 02 2015 DIQ S 181 UNIVERSAL TRANSMITTER FOR ONE DIGITAL SENSOR...
Страница 2: ...Copyright 2016 Xylem Analytics Germany GmbH Printed in Germany...
Страница 6: ...List of contents DIQ S 181 6 ba77083e01 02 2015...
Страница 102: ...Index DIQ S 181 102 ba77083e01 02 2015...
Страница 103: ......