Cond 1970i
Operation
19
ba75692e02
07/2009
4.2.4
TDS (Total dissolved solids)
You can measure the total dissolved solids as follows:
1
Perform the preparatory activities according to section 4.2.1.
2
Immerse the conductivity measuring cell in the test sample.
3
Press the
<M>
key until the unit mg/l
appears.
The value of the total dissolved solids appears in the upper dis-
play line. The TDS factor appears in the lower display line.
4
Using
<
▲
> <
▼
>
, set the TDS factor (0.40 ... 1.00).
(The TDS factor has to be determined by a comparative mea-
surement before.)
5
Wait for a stable measured value.
TP
°
C
Sal
%
REL1
ARng
S
C
cm
S/
Sal
Cal
RCL
%
cm
1/
0
m
O
Oxi
AR
Year
Ident
No.
Day.Month
Tref20
LoBat
Lin
pH
TEC
Auto
Store
Time
Baud
K
/
REL2
pH
/
mV
1
0 0
4
4 7
mg/ l
Tref25
nLF
TP
°
C
Sal
%
REL1
ARng
S
C
cm
S/
Sal
Cal
RCL
%
cm
1/
0
m
O
Oxi
AR
Year
Ident
No.
Day.Month
Tref20
LoBat
Lin
pH
TEC
Auto
Store
Time
Baud
K
/
REL2
pH
/
mV
0
8
2
8
3
9
mg/ l
Tref25
nLF
Содержание Cond 1970i
Страница 44: ...Operation Cond 1970i 44 ba75692e02 07 2009 TDS Measuring range Voltage Resolution 0 1999 mg l 0 1999 mV 1 mV...
Страница 57: ...Cond 1970i Technical data 57 ba75692e02 07 2009 Climatic class VDI VDE 3540 IP protection EN 60529 1991...
Страница 60: ...Technical data Cond 1970i 60 ba75692e02 07 2009...
Страница 66: ...Index Cond 1970i 66 ba75692e02 07 2009...