pH 298, LF 298, Oxi 298, Cl 298
Communication with Modbus RTU
55
ba77196e03
07/2018
6
Communication with Modbus RTU
6.1
Connecting the Modbus interfaces
The Modbus is connected to the Modbus connectors (X3.4 ... X3.6) of the
measuring transmitter.
General instal-
lation instructions
Always install the wires with a minimum distance of 20 cm from any other wires
carrying a voltage of more than 60 V.
Tools
Phillips screwdriver
Small screwdriver.
NOTE
Modbus cables may only be connected to the Modbus connectors (X3.4 ...
X3.6). No wire of the cable may be connected with an external electrical poten-
tial. Otherwise, malfunctions could occur.
1
Remove the cover on the left and right-hand side of the measuring
transmitter.
2
Remove two cross-head screws each on the left and right-hand side of
the measuring transmitter.
3
Tilt down the front of the enclosure.
4
Loosen the coupling ring of the cable gland at the position for the
Modbus connection (X3.4 ... 6) on the enclosure.
5
Feed the Modbus cable into the enclosure through the cable gland.