LWCC-M-10

MicroLWCC (WPI #

LWCC-M-10

) is a new fiber optic, low volume flow cell for UV/

VIS/NIR absorbance analysis. 

Based on WPI’s established 

liquid core waveguide tech-

nology, the analyte solution 

functions as the core of a fluid 

filled light waveguide. Wetted 

parts in the sample cell light 

path are PEEK, fused silica and 

Teflon. Optical fibers are used 

to transport light to and from 

the sample cell. The cell can be 

used in biochemistry for DNA, 

RNA and protein quantification, 

colorimetric nutrient and trace 

metal analysis, drug discovery 

and dissolution testing, process 

control and HPLC analysis.

ASSeMbLy

Light source and detector can be connected to the 

LWCC-M-10

 via two SMA-termi-

nated fiber optic cables with a core diameter of 600µm. The fiber optic connections 

are interchangeable in that either connector can be used to connect to the light 

source or to the spectrometer.  
As with the fiber optic connectors, the liquid ports are also interchangeable. It makes 

no difference which port is set-up as inflow and which port serves as the outflow. 

However, if an in-line filter is to be installed, it should be connected on the input side 

of the liquid flow path. Use standard HPLC type 10-32 coned PEEK 1/16

th

 tubing fit-

tings (WPI #

800827

) to make connections to the pump and waste receptacle.

OpeRATINg INSTRUCTIONS

The following procedures can be used to verify the transmission, noise and refractive 

index of your 

LWCC-M

.

Transmission Verification

1.  Connect one 600

m

m diameter fiber optic assembly to a light source and one 

600

m

m diameter fiber optic assembly to the detector.

2.  Use an SMA bulkhead adapter (WPI #

13395

) to butt-couple the fibers. This is the 

reference for the measurement of transmission.

3.  Remove the SMA bulkhead adapter and insert the 

LWCC-M

4.  Record the percent transmission. Values for the 254nm, 280nm and 550nm 

should exceed 40%.

Noise Determination

Noise is determined using the ASTME685-93 standard as reapproved in 2000. Use 

methanol as the solvent at a flow rate of 1.0mL/min. Using the WPI test setup, noise 

values are below 35uAU. See www.astm.org/Standards/E685.htm to obtain a copy of 

the standards document.

Refractive Index

The refractive index between methanol and cyclohexane is measured using a flow 

rate of 3.0mL/min. Because of the immiscibility of the solvents, ethanol is used be-

tween the solvents to prevent unwanted interaction between methanol and clyclohex-

ane. An RI value between zero and seven is acceptable at the 280nm wavelength.

MAINTeNANCe

Cleaning

With normal use, the 

LWCC-M-10

 will rarely need cleaning. If measurement results 

diminish, clean your 

LWCC-M-10

. Use the cleaning kit for liquid waveguides (WPI 

#

501609

).

1.   Connect the exit tubing (for example, silicon or equivalent) from the fluid output 

of the 

LWCC-M-10

 to a waste container.

2.  Rinse the cell thoroughly using Ultra Pure water. Obtain a new reference intensity 

and take a baseline absorbance reading.

3.  Inject 1cc “Cleaning Solution 1” into the 

LWCC-M-10

 fluid input port.

4.  Inject 1cc “Methanol Solution 2” into the fluid input port of the 

LWCC-M-10

.

5.   Inject 1cc “HCl Solution 3” into the fluid input port of the 

LWCC-M-10

.

6.  Flush out the cleaning solutions with distilled, Ultra Pure, reverse osmosis or 

equivalent quality water and take an absorbance reading.

7.  Repeat these cleaning cycles until a stable absorbance signal can be obtained.

8.   If the UV readings are still low after cleaning, swab the fiber optic surfaces of the 

sample cell with a fiber optic Q-tip dipped in isopropyl alcohol.

Storage

To store the instrument, clean the 

LWCC-M-10

, and then seal the fluid input and 

output ports using either the caps provided or an alternative.

SpeCIfICATIONS

LWCC-M-10

Optical pathlength

10mm

Internal volume

2.4

m

L

Wavelength Range

200–1000nm

Fiber Connection

(SMA) 600µm

Transmission at 254nm*

> 40%

Liquid Connection

10-32 coned port for 1/16

th

 tubing

Maximum Pressure

1000 PSI

Refractive Index at 280nm**

< 7 mAU

Wetted materials

PEEK, Fused Silica, Teflon

* Reference: Two 600µm fibers, butt-coupled

** Measured using ASTM E 685 - 93

WPI U.S. Patents: 5,444,807; 5,570,447; 5,604,587; 6,603,556; 6,385,380.

Light Input

600

µ

m S

MA

Light Output

600

µ

m S

MA

Fluid Input

800827 Connector

Fluid Output

800827 Connector

1.00"

1.3

8”

Fig. 1–LWCC-M-10 showing connections. 