Setting up TH Thermometer Amplifier and Sensor

The TH amplifier and its sensor are designed to measure the temperature of the fluid 
perfusing a heart or an organ (liver, kidney, mesentery) in order to determine the 
temperature of the organ. 

 

Fig. 6—The TH Temperature Amplifier, front and back.

If the temperature sensor is going to be used:
1.  Plug the microphone jack on the end of the temperature sensor to the 

TH IN

 

socket on the back panel of the amplifier. The sensor is now connected to the 
TH amplifier module.

2.  Plug the microphone jack of one of the output cables into the 

TH OUT

 socket of 

the TH pressure amplifier. Connect the other end of the output cable to one of 
the analog inputs of the data acquisition unit.

3.  Turn the 3-way program switch to 

GND

. Place the temperature probe on or into 

the area where the temperature needs to be measured. Turn on the amplifier 
unit. The actual temperature of the area being monitored appears on the 3-digit 
display on the front panel of the TH module.

4.  Check the baseline level of the temperature recording channel. A signal of 

zero (0) volts is sent by the amplifier to the data recorder. If necessary, set the 
baseline of the recording channel to zero (0) using the controls of the data 
recording system.

5.  Turn the 3-way program switch to the 

45ºC

 position. The voltage level on the 

recording channel should change from 0 to 450mVDC corresponding to the 
programmed 0-45ºC temperature range.

6.  Turn the program switch to 

MEAS

. The voltage level on the recording channel 

should now correspond to the actual temperature at the location of the sensor.

Содержание Biological Signals Amplifier

Страница 1: ...Biological Signals Amplifier for isolated heart Langendorff Neely and perfused organ systems World Precision Instruments INSTRUCTION MANUAL Serial No _____________________ 070212 www wpiinc com...

Страница 2: ......

Страница 3: ...prior written permission of World Precision Instruments Inc CONTENTS ABOUT THIS MANUAL 1 INTRODUCTION 1 Parts List 2 Unpacking 2 INSTRUMENT DESCRIPTION 3 Setting Up HG Pressure Amplifiers and Sensors...

Страница 4: ...iv World Precision Instruments...

Страница 5: ...y perfusate pressures temperature and ECG signals or surface potentials from isolated perfused heart Langendorff and working heart Neely preparations The system can also be used to measure perfusion p...

Страница 6: ...at sales wpiinc com or 941 371 1003 to speak with one of our representatives regarding the type of electrodes needed to record the biopotentials of interest Unpacking Upon receipt of this instrument m...

Страница 7: ...ls Amplifier World Precision Instruments 3 INSTRUMENT DESCRIPTION Fig 2 The front panel has the operation switches and indicators Fig 3 The back panel has the input and output sockets along with the m...

Страница 8: ...cket of the HG1 pressure amplifier Connect the other end of the output cable to one of the analog inputs of the data acquisition unit 3 Before the transducer is connected to a perfusion cannula or cat...

Страница 9: ...pressure module as described in the previous section 2 Turn the 4 way program switch on the front panel of HG amplifier to GND Set the BALANCE potentiometer to the middle position Turn on the amplifi...

Страница 10: ...log inputs of the data acquisition unit 3 Turn the 3 way program switch to GND Place the temperature probe on or into the area where the temperature needs to be measured Turn on the amplifier unit The...

Страница 11: ...ler fluid Place the tip of the probe as close to the bulb of the thermometer as possible without touching it 4 Turn the 3 way program switch to the MEAS position Wait until the temperature reading on...

Страница 12: ...of interest The type of electrodes plate pellet needle wire etc determine the type of lead wires needed to connect the electrodes to the input cable of the ECG amplifier and the preparation required...

Страница 13: ...ding channel Turn the 3 way program switch to 1mV The signal on the recording channel should show a deflection of 1mVDC which is the nominal voltage used to calibrate ECG amplifiers If needed use the...

Страница 14: ...10 World Precision Instruments...

Страница 15: ...ct indirect special or consequential arising from a failure of this product to operate in the manner desired by the user WPI shall not be liable for any damage to data or property that may be caused d...

Страница 16: ...44 0 1462 424700 Fax 44 0 1462 424701 E mail wpiuk wpi europe com Germany Zossener Str 55 10961 Berlin Tel 030 6188845 Fax 030 6188670 E mail wpide wpi europe com China Hong Kong WPI Shanghai Trading...

Отзывы: