background image

Calibration Unit 

for PGT

®

120 

Calibration manual for PGT120 with DIP12 switch 

Содержание 7100.PGT120.CU

Страница 1: ...Calibration Unit for PGT 120 Calibration manual for PGT120 with DIP12 switch...

Страница 2: ...4 OFF 8 3 2 1 2 Wrist strap verification To check the limits of the wrist strap test connect the middle jack of the calibration unit with the 3 mm snap of the PGT 120 same symbols Use the DK3 socket a...

Страница 3: ...h the red 4 mm socket on the rear side of the PGT 120 footwear electrode same symbols Set the marked lever of the rotary switch in succession to the positions mentioned below Press the right contact e...

Страница 4: ...the PGT 120 footwear electrode same symbols 4 1 DIP switch setting for upper limit 35 M Switch 3 Switch 4 ON 5 4 OFF ON OFF 8 3 2 1 Set the marked lever of the rotary switch in succession to the posit...

Страница 5: ...upper limit 140 M for series Switch 8 ON 8 5 3 ON OFF 4 2 1 Set the marked lever of the rotary switch in succession to the positions mentioned below Reset the instrument after each measurement by dis...

Страница 6: ...e Calibration Unit and set the marked lever of the rotary switch in succession to the marked positions The corresponding resistor values and tolerances can be taken from the drawing below Also connect...

Страница 7: ...S directive We identify our electrical and electronic devices as of August 2005 in accordance with WEEE 2012 19 EU and ElektroG with the symbol shown to the right per DIN EN 50419 These devices may no...