Save measurements- M00 Memory-Mode
43
EN
The bar color changes depending on the differ-
ence between the infrared temperature IRTand
the dew point temperature DPT
IRT – DPT
> 1,9° °C: yellow
1,1 °C <IRT – DPT
≤
1,9 °C: blue +
Icon
- 0,5 °C < IRT – DPT
≤
1,1 °C: blue +
Icon
IRT – DPT
≤
- 0,5 °C: blue +
Icon
HINWEIS!
The drop icon is the indication for the risk of con-
densation as well as mold warning. The buzzer
and optional vibration functions are activated at
the same time.
13 Save measurements- M00 Memory-Mode
Save measurements
It is possible to store up to 50 data sets in the
device and to display them.
•
Measure the temperature value.
•
Press the Mode button repeatedly until the
memory mode M00 is called up.
•
Press the measuring button to save the cur-
rently displayed data set as M01.
NOTE!
After switching the device off and on again, the
stored data records are retained.
Display stored data
•
In memory mode M00, press the up/down key
until the desired data set is displayed.
Delete records
•
To delete all records, press the Mode button
to enter Momory mode M00.
•
Press and hold the Up button. Now press the
down button simultaneously for 2 seconds.
Clr flashes on the display and the measured data
is deleted.