www.wobit.com.pl
Manual -MD150T – 27.10.2014 – v.1.3
page 14
Rated load
for sensor it is a size of force, which specifies
top value of measuring range (given value for sensor as
bridge range
-
TRA
in N).
Applied load
is the biggest force, which have
unambiguous connection between force and output
signal (signal value is changing linear in relation to force). It can count max. M
150%
of sensor range
(bridge).
Cut-off load
is the biggest force, which can Take each sensor without it damage. It can count max.
200%
of sensor range.
Destroying load
means force applied to sensor Shaft, which exceeding can cause to mechanical
damage of the sensor.
Constant of e.g. KMM60 sensor is 1,5mV/V
2%. Error of the sensor is specified by following values:
tolerance of linearity 0,2% of measuring range, zero tolerance 2% of measuring range, hysteresis
0,2% of range and creep measured in time of 30 minutes 0,1%. Temperature factor of zero point
and constant is 0,1% of range/10
o
C. Bridge resistance on input has 380om ±10% and output
resistance has 350om ±5%.
9.
Technical parameters
Description
Parameter
Power supply
12 … 36 VDC, 250mA
Sensor input
Sensor power supply: 5V
Max. differential voltage: ±39mV,
Resolution: 0,001% FS,
Temperature error: 0,0025%/C°
Frequency of measurements: 10Hz, 80Hz
Zeroing input TARA
Low level: 0V (max. 2V), high level: +24V (5…24V)
Minimal pulse length >10ms, minimal pause. 0,5sek.
Relay outputs PK1, PK2
2 x 1A/125VAC, 2A/30VDC
Interfaces
RS485 MODBUS-RTU
,
default parameters 38400bps, 8:n:1,
USB: 1.1, 2.0
Operating temperature range
0..50
o
C
Display
6 digits, height13.5 mm
Housing
height: 45 mm
width: 92 mm
length: 81 mm
Weight
200g
Degree of protection
IP40, from panel head – IP65
Universal password
3145
Applied load
-150% TRA
Cut-off load
-200% TRA
Destroying load
<200% TRA