The inoculation loop sterilizer Sterimax is ideal for laboratory use for
sterilizing inoculation loops, tweezers and micro-instruments made of
heat resistant materials such as stainless steel or platinum-iridium alloys.
General safety advise
!
The range
SteriMax - Inoculation Loop Sterilizer:
Length 118 mm, Ø 15 mm:
Length 161 mm, Ø 20 mm:
230 V / AC
Art. No.: 5.001.000
230 V / AC
Art. Nr.: 5.101.000
120 V / AC
Art. No.: 5.002.000
120 V / AC
Art. Nr.: 5.102.000
100 V / AC
Art. No.: 5.003.000
100 V / AC
Art. Nr.: 5.103.000
•
with 7 angular adjustments
•
1 screwdriver (R1)
•
1 inoculation loop holder (R2)
•
1 inoculation loop Ø 3 mm (R3)
•
Tray for 3 WLD-TEC loop holder (13)
•
Power cable (11)
•
2-year warranty
11
R3
R2
R1
Fig. F
13
Only operate the annealing system under constant supervision!
General note
Ceramic insulated heating element: The inital brown color of the heating
element insulation is normal. The heating element decolors during use.
i
Clean the ceramic annealing tube frequently and
check for damage. If the ceramic annealing tube
(3b) exhibits any cracks, chips or similar, do not
use the device! Replace the ceramic annealing
tube. Check also if the quartz glass heating
element (3a) is damaged! For further information
see paragraph 6 of this manual.
3a
3b