11.4 Pump
faulty
The device does not "hum" once the measurement has been started.
The first thing to do is rule out other causes.
Check whether the needle or filter are clogged up (see Chapter 11.1).
Check whether a hose has been kinked or has been crushed.
Switch off the device.
If plugged in:
detach USB cable or USB mains cable.
On the rear side of the OXYBABY
®
:
unscrew the 6 fastening screws and open the casing.
Check hoses, if kinked or defective, send the OXYBABY
®
to
the WITT after-sales service.
After checking:
close the casing and screw back together again using 6 fastening screws.
If no other cause can be determined, send the OXYBABY
®
to the WITT after-sales service.