background image

For the right use and safety of this product, please read through this manual prior to use.

Handling error may cause device trip, injury, or disaster.

1. For the safety, fixing work shall be done by the authorized with instrument or electrical 

2. Use the device within the rated in/out value as written in the specification.
   Otherwise, the device may fall in trouble.

3. Fix this product to the place under the environment written in the specification.
   Otherwise, it may fall in trouble.

4. As for wiring, follow the internal wiring rules and the electric facility technology standard.

5. Be sure to shut off the power before wiring.
   Otherwise, the device may fall in trouble.

6. Both ends of a cable shall be the solderless terminal coated with an insulator.

7. Do not disassemble this product at all costs.
   Otherwise, it may fall in trouble.

8. Use a proper protection tube to fix this product to a tube.

9. Select the protection tube according to the process condition.
   Please remember the owner is liable for wrong protection tube selection.

10. A head o-ring subject to aged deterioration or change may lose its sealing performance, 
     so it must be exchanged regularly, preferably every 5 years.

Instructions for proper and safe operation

WARNING

Содержание R100 Series

Страница 1: ...USER MANUAL PRODUCT NAME TEMPERATURE SENSOR MODEL R100 R200 R300 R400 R600 R700 R900 SERIES ...

Страница 2: ...nal wiring rules and the electric facility technology standard 5 Be sure to shut off the power before wiring Otherwise the device may fall in trouble 6 Both ends of a cable shall be the solderless terminal coated with an insulator 7 Do not disassemble this product at all costs Otherwise it may fall in trouble 8 Use a proper protection tube to fix this product to a tube 9 Select the protection tube...

Страница 3: ...sibility 6 Product Refund and Packing 7 Maintenance 8 Fault 9 Appearance and Function 10 Temperature Sensor Series 10 1 Head Terminal Box 10 2 Connection Part Type 11 Protection Tube 12 Temperature Sensor Protection Tube Installation Position 13 Connection CONTENTS ...

Страница 4: ...tection tube and you may select a protection tube according to the progress 3 Warranty We are not liable for any damage or failure of this product which was caused by your own modification change or repair against this manual and the warranty will be no longer valid 4 Nameplate 4 1 IEC Explosionproof Grade 1 Type 4 Tag No 2 Explosionproof Grade and Allowable Temp 5 Warning Text 3 Explosionproof Ce...

Страница 5: ...corrosive abrasive fluids or fluids with the flow rate and pressure WISE may support the user s selection but is not responsible for such selection II 2G Ex d IIC T6 Gb Tamb 20 C to 60 C LCIE 06 ATEX 6072X 02 Tag No TYPE ETR Series TYPE ETR10 Series KEMA 09ATEX0006X II 1 2 G Ex ia IIC T6 T1 Ga Gb Tamb 20 C to 60 C Ui 30V Ii 100mA Pi see the instruction manual Ci 0 1nF Li 0 01mH Tag No 0344 199 San...

Страница 6: ...ge during transportation shall be recorded by writing and all the losses from installation delay may be charged upon the transportation company for compensation 7 Maintenance 7 1 Maintenance In general temperature sensors are maintenance free If you need to perform maintanance activities maintanance may only be performed by skilled personnel and WISE CONTROL can assist in maintenance 7 2 Repair Re...

Страница 7: ... and Function 9 1 Temperature Sensor Appearance 9 2 Measuring Insert Appearance A Measuring Insert O D B Inorganic Insulating Sheath MgO Filled C Temp Sensor Length D Unbendable Length 100 M Measuring Insert Length ...

Страница 8: ...Low The temperature measuring element is a temperature sensor under EN60584 thermocouple or EN60751 thermoresistor Standard The temperature sensor may be manufactured either with or without the transmitter 10 Temperature Sensor Series 10 1 Head Terminal Box 10 2 Connection Part Type 10 2 1 General Head Small Type General Type Explosionproof Explosionproof 2Ports ...

Страница 9: ...10 2 2 Explosionproof Head 10 2 3 Relation between Connection Length and Head Temp 800 620 430 250 T ...

Страница 10: ...n tube shall be selected by the user customer according to the process Of course we can give you advice to help your selection but we are not liable for it 11 2 Protection Tube Type 11 2 1 By Connection Methods 1 Thread type 2 Flange type 3 Welding type 11 2 2 By Materials 1 End Close Type Where there is neither fluid flow nor pressure 2 Drilled Bar Type 11 2 3 By Protection Tube Appearances 1 Str...

Страница 11: ...11 3 Protection Tube Layout ...

Страница 12: ...line for installation 5 Minimize the effect of the natural frequency and make the inserting length as short as possible 50 60 6 In case of screw type connector use a sealant for connection 7 To install the sensor at a place having fluid flow make the temperature sensing part against the flow If infeasible taper it or make a right angle for installation in worst cases 12 2 Attachment Condition a To...

Страница 13: ...b To Straight Tube c To T Confluent Tube Taper Right Angle Taper Right Angle Taper Right Angle ...

Страница 14: ...d To Decreasing Tube e To Other Device f To Header Use inserting length less than 300 Taper Right Angle Taper Right Angle ...

Страница 15: ... power prior to explosionproof grade product connection 2 Be sure to check the polarity of the thermocouple prior to connection and discriminate terminal A B B for three wire for thermoresistor 3 If necessary connect a ground wire to the head 4 Use a solderless terminal to attach the extension wire or cable end to the terminal block and fasten it firmly with screws 5 In case of explosionproof grad...

Страница 16: ...3 3 Temperature Sensor Terminal Connection 13 3 1 thermoresistor 13 3 2 Thermocouple General Explosionpro SINGLE 3 WIRE DOUBLE 6 WIRE General Head SINGLE 2 WIRE DOUBLE 4 WIRE SINGLE 3 WIRE DOUBLE 6 WIRE ...

Страница 17: ...13 3 3 Terminal Box and Fastening Screw Size General 50 40 M4 0 7P Screw Size Connection Screw Pitch 60 5 M3 0 5P O D 67 5 Explosionproof Explosionpro ...