If it is necessary to replace any parts due to wear or damage, Winncare Nordic ApS 
recommends for safety reasons to use only original spare parts. Use of non-original spare parts 
will void warranty. 

 

Warranty 

There is a 5-year warranty on the product conditional on annual service and maintenance being 
carried out, according to the above service guidelines. See further warranty conditions on 
www.ergolet.com

 

 

4.2

 

Troubleshooting 

 

 

Fault: 

Possible cause 

Solution 

The Luna Mover does not 
react when the hand 
control LEFT/RIGHT are 
used. 
 

Batteries are not fully 
charged or defective. 

Charge the Luna ceiling hoist as 
described in Luna user manual 

The emergency stop is 
activated on the Luna and 
the red light is on. 

Deactivate emergency stop. Hold 

“R” button

 on the hand control for 

5 seconds. 

The power cable is not 
properly connected. 

Attach power cable as seen in 2.1 

Defective hand control. 

Check if up/down buttons on the 
hand control works. If not try a 
different hand control or order a 
new from the dealer or 

 

Winncare 

Nordic ApS . 

Defective Luna 

Check the chapter 

“troubleshooting” in the Luna user 

manual. 

Fault: 

Possible cause 

Solution 

The Luna Movers motor is 
running but it is not 
moving. 

The Luna Mover is 
unlocked. 

Use the safety release handle to 
lock the Luna Mover (See 2.2). 

Содержание Luna Mover

Страница 1: ...ion Please read this user manual carefully before using the product This user manual is intended as a guide for caregivers and other users as well as technical personnel during servicing and maintenan...

Страница 2: ...2...

Страница 3: ...n of safety functions ____________________7 3 MAINTENANCE 3 1 Cleaning _____________________________________8 3 2 Storage ______________________________________8 4 CHECK WARRANTY AND TROUBLESHOOTING 4...

Страница 4: ...ions before using the product NEVER use the Luna Mover if there are any indications of wear or suspicion of other defects on the track systems the Luna Mover hoist or on the accessories In such cases...

Страница 5: ...ckers for direction indication 1 6 Transport If the Luna Mover is to be transported between sites it is recommended to use the original packaging In case of return shipment of the Luna Mover the origi...

Страница 6: ...heck that the safety release handle is not activated If done correctly the Luna Mover cannot be moved by hand See 2 2 2 Lift up the ceiling hoist A and connect to the Luna Mover resistance will be fel...

Страница 7: ...ety release handle In case of a Luna Mover malfunction it will not be able to move Should a patient be suspended during a malfunction the safety release handle can be used to unlock the Luna Mover mov...

Страница 8: ...qualified technician at least once a year in accordance to Winncare Nordic ApS service instructions An annual check is also recommended to the European Standard EN ISO 10535 for patient lifters hoist...

Страница 9: ...teries are not fully charged or defective Charge the Luna ceiling hoist as described in Luna user manual The emergency stop is activated on the Luna and the red light is on Deactivate emergency stop H...

Страница 10: ...easurements V In Covering Noise IP class 275 kg 0 25 m s 2 9 kg 270 x 180 x 200 mm 24 V High impact moulded ABS During transfer without load 60dB A During transfer with 275 kg 55dB A IPx4 5 2 Dimensio...

Страница 11: ...he environment if not handled in an environmentally responsible way 6 2 Quality Winncare Nordic ApS is certified according to the quality standard ISO 9001 and ISO 13485 These certifications mean that...

Страница 12: ...S Taarnborgvej 12 C 4220 Korsoer Denmark Phone 45 70 27 37 20 Fax 45 70 27 37 19 Email info ergolet dk www ergolet com 93 42 E F CLASS I EN ISO 10535 EN 60601 1 2 UM 910 00784UK SWL 275 kg Scan the QR...

Отзывы: