Winncare Luna Mover Скачать руководство пользователя страница 8

 

3

 

MAINTENANCE 

3.1

 

Cleaning 

The Luna Mover should be cleaned using a slightly damp cloth with a small quantity of standard 
household cleaning agent or use surgical spirit to over-drying. 

Strong chemicals or abrasive agents must not be used.

 

3.2

 

Storage 

The Mover must be stored in a dry environment with relative humidity under 70%. 

4

 

CHECK, WARRANTY AND TROUBLESHOOTING 

4.1

 

Annual check and warranty 

The Luna Mover must always be repaired and maintained by qualified personnel. 

The manufacturer Winncare Nordic ApS recommends that the Luna Mover MUST be checked 
by a qualified technician at least once a year in accordance to Winncare Nordic ApS service 
instructions. An annual check is also recommended to the European Standard (EN/ISO 10535) 
for patient lifters/hoists including ceiling hoist. 

It is the legal owner’s

 duty to ensure that these checks are carried out on the product. 

The following items is to be inspected during the annual check: 

 

Cart wheels and bearings 

o

 

Check for noise and change if necessary  

 

Driving wheels/Rubber discs 

o

 

Clean for dust and check for wear 

 

Bearings for driving wheel 

o

 

Check if worn down, change if necessary 

 

Wire and motor connection 

o

 

Check for defect 

 

Safety release system 

o

 

Check for defect 

 

Overall condition 

o

 

Cracks, wear and tear, etc.  

Note!

 Scan the QR-code and see the instructions on video.

 

 

Содержание Luna Mover

Страница 1: ...ion Please read this user manual carefully before using the product This user manual is intended as a guide for caregivers and other users as well as technical personnel during servicing and maintenan...

Страница 2: ...2...

Страница 3: ...n of safety functions ____________________7 3 MAINTENANCE 3 1 Cleaning _____________________________________8 3 2 Storage ______________________________________8 4 CHECK WARRANTY AND TROUBLESHOOTING 4...

Страница 4: ...ions before using the product NEVER use the Luna Mover if there are any indications of wear or suspicion of other defects on the track systems the Luna Mover hoist or on the accessories In such cases...

Страница 5: ...ckers for direction indication 1 6 Transport If the Luna Mover is to be transported between sites it is recommended to use the original packaging In case of return shipment of the Luna Mover the origi...

Страница 6: ...heck that the safety release handle is not activated If done correctly the Luna Mover cannot be moved by hand See 2 2 2 Lift up the ceiling hoist A and connect to the Luna Mover resistance will be fel...

Страница 7: ...ety release handle In case of a Luna Mover malfunction it will not be able to move Should a patient be suspended during a malfunction the safety release handle can be used to unlock the Luna Mover mov...

Страница 8: ...qualified technician at least once a year in accordance to Winncare Nordic ApS service instructions An annual check is also recommended to the European Standard EN ISO 10535 for patient lifters hoist...

Страница 9: ...teries are not fully charged or defective Charge the Luna ceiling hoist as described in Luna user manual The emergency stop is activated on the Luna and the red light is on Deactivate emergency stop H...