6
Side foil:
(10)
-
Plug connection for hand control.
(11)
-
Data plate
–
Information of serial
number and symbols.
2.2
Safety functions
Power outage:
In case of a power failure the ErgoTrainer brakes will activate, so it cannot be moved in the track
(See 3.2)
Note!
The lifting function
UP/DOWN
still works.
Emergency stop:
This function is only to be used in an emergency e.g., if the lift function does not stop after
releasing UP/DOWN button. The emergency STOP button is placed on the hand control and on
the side foil.
If the emergency STOP is activated all functions will stop immediately and the red LED will be lit
on the hand control.
The emergency STOP can be deactivated by pressing the Reset button on the hand control for 5
seconds.
Emergency lowering button:
The emergency lowering feature is only to be used in an emergency where it is necessary to
safely lower the user. The emergency lowering button is placed on the hand control, illustrated
as a parachute and on the Luna side foil with 3 arrows.
To activate the emergency lowering press and hold the button on the hand control or on the side
panel for 5 seconds
–
“fast continuous beeping sounds”.
Warning!
The emergency lowering does not work if the hand control is defect.
Note!
When using the emergency lowering feature the strap will not stop at the lower end stop!
Release the emergency lowering button before the strap is at its maximum length.
Note!
If the emergency stop functions has been activated due to an emergency the
ErgoTrainer must not be used before a check has been made and any defect repaired.
Note!
If questions, please contact Winncare or your dealer.