WMP-24C/24F User’s manual
9
Guidance and manufacturer’s declaration – electromagnetic
immunity
The model WMP-24C/24F is intended for use in the electromagnetic
environment specified below. The customer or the user of the model
WMP-24C/24F should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic
environment
–
guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
8 kV contact
15 kV air
8 kV contact
15 kV air
Floors should be
wood, concrete or
ceramic tile. If floors
are covered with
synthetic material, the
relative humidity
should be at least 30
%.
Electrical fast
transient/burst
IEC 61000-4-4
2 kV for
power
supply lines
1 kV for
input/output
lines
2 kV for
power
supply lines
1 kV for
input/output
lines
Mains power quality
should be that of a
typical commercial or
hospital environment.
Surge
IEC 61000-4-5
1 kV line(s)
to line(s)
2 kV line(s)
to earth
1 kV line(s)
to line(s)
2 kV line(s)
to earth
Mains power quality
should be that of a
typical commercial or
hospital environment.
Содержание WMP-24C
Страница 27: ...WMP 24C 24F User s manual 14 Dimension WMP 24F VESA Mount Screw type M4...
Страница 28: ...WMP 24C 24F User s manual 15...
Страница 29: ...WMP 24C 24F User s manual 16...
Страница 30: ...WMP 24C 24F User s manual 17 WMP 24C VESA Mount Screw type M4...
Страница 31: ...WMP 24C 24F User s manual 18...
Страница 32: ...WMP 24C 24F User s manual 19...
Страница 63: ...WMP 24C 24F User s manual 50 J13 J15 DIMM1 DIMM0 DDR4 Vertical Connect...