Document No.: TMM-2217-13
Revision: A
Page 28 of 30
BATTERY SAFETY INFORMATION
TO PREVENT PERSONAL INJURY TO PATIENT AND / OR MEDICAL PERSONNEL, ENSURE
BATTERY REPLACEMENT IS PROPERLY INSTALLED AND SECURELY ENGAGED PRIOR TO USE OF
ACCESSORY.
TO PREVENT PERSONAL INJURY TO PATIENT AND / OR MEDICAL PERSONNEL, ENSURE
BATTERY REPLACEMENT IS INSTALLED BY PERSONNEL WHO HAVE THOROUGHLY READ INSTALLATION
INSTRUCTIONS.
PROPERLY DISPOSE OF BATTERIES BEING REPLACED DURING SERVICING
•
Battery posts, terminals, and related components contain lead and lead compounds, which are known by the
state of California to cause cancer or other reproductive harm. Wash hands thoroughly after handling or
servicing.
•
Lead battery is classified as “dangerous waste” and service technician is obliged by law to arrange for proper
disposal of battery, or recycling, to avoid battery refuse in environment.
•
Battery contains toxic material (lead) and corrosive fluid (sulfuric acid). Wear proper skin and eye protection
prior to handling battery. After exposure, wash skin thoroughly with water.
•
Do NOT short-circuit battery terminals, which can cause battery explosion or fire.
•
To request Material Safety Data Sheet (MSDS) for battery, contact TransMotion by Winco Customer Service.
BATTERY LIFE
•
Battery capacity diminishes over time. Charge batteries regularly to maintain healthy battery life.
•
Recommended to replace battery every two years or as required.
Suggested Protocol for Charging Your TMM Battery:
1. Ensure the cord for your TMM Battery Charger is fully inserted into the battery charger inlet.
2. Plug your TMM Battery Charger directly into an outlet. Do NOT use an extension cord or power strip. Extension
cords and/or power strips can cause a “drop” in current that may hinder the charger’s ability to charge the battery
efficiently.
3. Slide the TMM Battery downward onto the TMM Battery Charger making sure the battery is properly seated onto
the charging port.
4. Check to see if either an orange light (indicating the battery is “charging”) or a green light (indicating the battery is
“fully charged”) appears.
5. Charge your TMM Battery for eight (8) hours or until the Green light indicator appears to ensure an optimum charge
is achieved.
6. Rotate batteries from charger to chair every two (2) to three (3) days to maintain an optimum charge on the
batteries.
7. When your TMM Stretcher-Chair is not in use for extended periods of time (four (4) days or more) unplug the
battery from the Stretcher-Chair to prevent battery drainage.
8. Replace your TMM Battery every two (2) years (recommended) or as required.