Document No.: OM TMM4-X SERIES 

Revision: X 

Page 11 of 28 

CASTER BRAKE OPERATION 

BRAKE MODE

 

Activate braking system by pressing down on red end of either 
brake pedal located at base of chair. 

  

This mode prevents all four casters from swiveling and all 
wheels from spinning. 

 

NEUTRAL MODE  

Activate neutral mode by placing either brake pedal into a 
horizontal orientation. 

 

This mode allows all four casters to swivel and all wheels to 
spin freely. 

 

STEER MODE 

Activate steer-locking system by pressing down on green end 
of either brake pedal located at base of chair. 

  

This mode locks caster (near patient’s right foot) parallel to 
base, but allows this wheel to spin.  Other three casters 
swivel, and wheels spin freely. 
 

 

 

SIDE RAIL OPERATION 

LOWERING SIDE RAIL 

Grasp top of side rail and push inward slightly, while pulling out 
red rail release tab.  Once released, lower rail.

 

 

STOWING SIDE RAIL 

To stow side rail, lower rail into “down” position below seat (or 
back section depending on rail location).  Push side rail until it 
fully engages retainer clip. 

 

Содержание TransMotion Medical TMM4-X Series

Страница 1: ...el TMM4 X Series Multi Purpose Tilt Chair and Surgi Chair II Options W Wide T Tall R Rail on Back Section Q Scale L Low F Folding Footrest E Export B Battery A AC Power 5516 S W 1st Lane Ocala FL 3447...

Страница 2: ...ANGLE 12 PENDANT CONTROLLER 14 PATIENT INGRESS EGRESS 15 PATIENT TRANSFER 16 PUSH BAR OPERATION 16 OPTIONS 17 W OPTION WIDE WIDTH 17 T OPTION TALL HEIGHT 17 L OPTION LOW HEIGHT 18 R OPTION RAILS ON B...

Страница 3: ...isted are nominal and operation may vary slightly from unit to unit due to tolerance variations or power supply variations Photographs and illustrations contained within this document may not depict e...

Страница 4: ...osing battery to open flames cigarettes sparks and incandescent materials Never charge battery in enclosed unventilated spaces Do NOT store battery in sealed container Store in fresh well ventilated a...

Страница 5: ...se unpredictable operation resulting in injury to patient medical personnel or bystander Modifying chair will void warranty USE SAFE OPERATING PROCEDURES Prior to operating chair ensure patient cleara...

Страница 6: ...s likely EMI will interfere with chair operation The chair has been tested to 20 V m immunity level which is sufficient to provide useful protection from common sources of radiated electromagnetic ene...

Страница 7: ...re information For A Option Only If chair is to be stored or not in use for more than 1 5 weeks leave chair plugged into an AC wall outlet or unplug battery from control box For more information refer...

Страница 8: ...Document No OM TMM4 X SERIES Revision X Page 8 of 28...

Страница 9: ...Document No OM TMM4 X SERIES Revision X Page 9 of 28...

Страница 10: ...T SHOULD BE MOVED TO A MORE STABLE NON PADDED PLATFORM SECONDARILY IF ATTEMPTING TO PERFORM CHEST COMPRESSIONS ON THE TMM DEVICE THE PATIENT SHOULD BE PLACED ON A BACK BOARD OR POSITIONED ON THE CHAIR...

Страница 11: ...ll wheels to spin freely STEER MODE Activate steer locking system by pressing down on green end of either brake pedal located at base of chair This mode locks caster near patient s right foot parallel...

Страница 12: ...tically Ensure side rail is secure by pulling on rail after it is raised ADJUSTING SIDE RAIL ANGLE Remove cushion and seat pan by removing four pan head screws NOTICE Do not discard seat pan pan head...

Страница 13: ...nut to tighten top Once aligned top nut is tightening nut 2 When adjusting ensure bottom jam nut is contacting seat frame 3 Ball joint head angle does not matter 4 Prior to adjustment pull outward on...

Страница 14: ...section Press right button to lower chair s leg section SEAT HEIGHT ADJUSTMENT Third row of pendant buttons controls height of seat section Press left button to raise chair s seat Press right button t...

Страница 15: ...ster wheels 2 Adjust chair to lowest height and into upright chair position 3 Lower one side rail of seat section 4 If back section rails are present remove lower back section rail of same side 5 Posi...

Страница 16: ...into horizontal orientation Press Auto Contour button 2 Lower remove all side rails from transfer side of chair 3 Position chair as close as possible to other surface 4 Match chair stretcher height t...

Страница 17: ...OM TMM4 X SERIES Revision X Page 17 of 28 OPTIONS W OPTION WIDE WIDTH For W option patient surface width is 28 Standard width is 24 T OPTION TALL HEIGHT For T option chair height travel is 16 Standard...

Страница 18: ...Standard height is 25 Chair height travel is the same as Standard and T Option To prevent chair damage ensure ground clearance is at least 1 75 when driving chair up a ramp or over a bump R OPTION RA...

Страница 19: ...e measurement tare scale before each patient To prevent inaccuracy prevent extra items accessories monitors IV bags blankets pillows charts etc from being included excluded during weight checks To ens...

Страница 20: ...button is pressed 1 Remove battery cover on back of pendant by sliding it downwards 2 Take out battery and replace it with a new 9V Industrial battery 3 Replace battery cover To ensure proper scale f...

Страница 21: ...omfort stow footrest prior to articulating chair into supine stretcher position EXTENDED FOOTREST STOWED FOOTREST E OPTION EXPORT E option designates chair was shipped with power cord specific to rece...

Страница 22: ...CHAIR Place two round cam locks on back of battery into square openings of mounting bracket Once in place slide battery down until cam locks engage ADDITIONAL BATTERY INFORMATION All chairs come with...

Страница 23: ...batteries overnight i e during off times or 24 hours prior to service When the chair is plugged in and charging the pendant will not illuminate even when buttons are pushed If chair is to be stored o...

Страница 24: ...13 COLUMN PLUG TMM4 X OR PLUGS TMM4 XT PLUG INTO PHONO TO DIN ADAPTER PLUGS THEN PLUG INTO PORTS 1 AND PORT 2 FOR TMM4 XT LONGEST WIRE FROM WIRING ASSEMBLY IS THE BACK SECTION ACTUATOR PLUGS INTO PORT...

Страница 25: ...bels Actuators and Battery Pack As required in accordance with facility policies Wipe with damp cloth and allow to air dry Facility approved detergents disinfectants water Clean exterior surfaces only...

Страница 26: ...ery should be cleaned on a regular basis with a damp cloth soaked in a mild soap and water solution Avoid harsh detergents or chemicals that could damage the finish of your product If disinfecting wit...

Страница 27: ...paint burrs rough edges Remove burrs rough edges Touch up paint Metal file Touch up paint To prevent personnel cuts burrs and rough edges should be removed once found All Fastened Joints Every three m...

Страница 28: ...ch can cause battery explosion or fire To request Material Safety Data Sheet MSDS for battery contact TransMotion Medical Customer Service BATTERY LIFE Battery capacity diminishes over time Charge bat...

Отзывы: