WILDLIFE 1001 Скачать руководство пользователя страница 5

IOP 019A.1 

Effective Date: 1/11/2017 

Page 5 of 11 

 

 

d. 

Dissolved oxygen sensor

 

1) 

Highlight 

DO

. Press 

enter

2) 

Highlight 

Enabled

 and press 

enter

 to enable (

) the DO port. 

3) 

Highlight 

Sensor Type

 and press 

enter

.  

4) 

Highlight 

Polarographic

 (default) and press 

enter

 to confirm. 

5) 

Highlight 

Membrane

 and press 

enter

.  

6) 

Highlight 

1.25 PE Yellow

 and press 

enter

 to confirm. 

7) 

Be sure that 

Local DO

 and 

LDS

 are not enabled (

). 

8) 

Press 

 to return to the previous menu. 

e. 

Conductivity

 

1) 

Highlight 

Conductivity

. Press 

enter

2) 

Highlight 

Enabled

 and press 

enter

 to disable it (

). 

3) 

Press 

 to return to the previous menu. 

e. 

Salinity

 

1) 

Press 

 to return to the previous menu. 

2) 

Highlight 

Salinity

. Press 

enter

3) 

Use the numeric entry screen to enter a value of 0.1 and then 

store the value. 

4) 

Press 

Esc

 when finished. 

 

3. 

Setup display 

a. 

Press 

Sensor

 button, highlight 

Display

, press 

enter

b. 

pH sensor

 

1) 

Highlight 

ISE1(pH)

. Press 

enter

2) 

Highlight 

pH

 and press 

enter

 to enable it (

). 

3) 

Highlight 

pH mV

 and press 

enter

 to enable it (

). 

4) 

Press 

 to return to the previous menu. 

c. 

Ammonium sensor 

1) 

Highlight 

ISE2(NH4)

. Press 

enter

2) 

Highlight 

NH4-N mg/l

 and press 

enter

 to enable it (

). 

3) 

Highlight 

NH4 mV

 and press 

enter

 to disable it (

). 

4) 

Highlight 

NH3-N mg/l

 and press 

enter

 to enable it (

). 

5) 

Press 

 to return to the previous menu. 

d. 

Dissolved oxygen sensor

 

1) 

Highlight 

DO

. Press 

enter

2) 

Using the 

 and 

 arrows to move between the lines and 

enter

 to toggle, enable (

DO mg/l

 and disable (

) DO %L 

and DO ppm. 

3) 

Press 

 to return to the previous menu. 

e. 

Barometer 

1) 

Highlight 

Barometer

. Press 

enter

2) 

Highlight 

None

 and press 

enter

3) 

Press 

 to return to the previous menu. 

f. 

Temperature

 

1) 

Highlight 

Temperature

. Press 

enter

2) 

Highlight 

°C

 and press 

enter

 to select it (

). 

3) 

Press 

 to return to the previous menu. 

 

C. 

Sensor Installation (This section can be skipped if not installing or replacing any sensors. 

1. 

Remove/unscrew the metal guard or calibration cup if installed. 

2. 

Remove the red rubber port caps if the instrument was being stored without 

Содержание 1001

Страница 1: ...ntrol Centre 1219 Queen Street East Sault Ste Marie Ontario P6A 2E5 Canada INSTRUMENT OPERATING PROCEDURE INSTRUMENT Multiparameter Water Quality Instrument MODEL Pro Plus Meter 6050000 Quatro Cable 605790 Probes temperature conductivity 5560 pH 1001 or 1001A DO 2003 ammonium 1004 MANUFACTURER Yellow Springs Instrument Co Inc YSI PRECAUTIONS POTENTIAL INTERFERENCES General Greatest accuracy is ach...

Страница 2: ...ulated from ammonium sensor measurements Exposure to the high ionic content of pH buffers can cause a significant but temporary drift in the ammonium sensors Therefore when calibrating the pH sensor remove ammonium sensor from the cable bulkhead and plug the port After pH calibration is complete replace the ISE sensors and proceed with ammonium calibration Alternatively the ammonium sensor may be ...

Страница 3: ... screen Use to calibrate all parameters except temperature File Opens File Menu from any screen Use to view data and GLP files set up site and folder lists and delete data Backlight Press to turn the instrument backlight on and off and to adjust the display contrast when pressed with the left or right arrow key Right Arrow Use to navigate right in alpha numeric entry screens Can be pressed simulta...

Страница 4: ... B Instrument software setup 1 Press the Power key to turn the instrument on The instrument will briefly display the splash screen with the YSI logo then go directly to the main run screen The first time the instrument is powered up or if the instrument has had a battery change with batteries removed for more than 2 minutes you will need to set the date and time a Highlight Date Time from the Syst...

Страница 5: ...ress enter 2 Highlight pH and press enter to enable it 3 Highlight pH mV and press enter to enable it 4 Press to return to the previous menu c Ammonium sensor 1 Highlight ISE2 NH4 Press enter 2 Highlight NH4 N mg l and press enter to enable it 3 Highlight NH4 mV and press enter to disable it 4 Highlight NH3 N mg l and press enter to enable it 5 Press to return to the previous menu d Dissolved oxyg...

Страница 6: ...etely without feeling any resistance Damage to your cable sensor may occur if you force the parts together e Once completely installed the nut will seat flat against the bulkhead At this point use the included tool to turn the nut an additional to turn DO NOT overtighten Refer to the sensor installation sheet that is included with the conductivity temperature sensor for detailed instructions 4 Ins...

Страница 7: ... clockwise to engage threads and finger tighten Do not use a tool This connection is waterproof Refer to the sensor installation sheet included with each sensor for detailed instructions 7 Install the 2003 DO sensor membrane as follows a Change membrane and oxygen probe electrolyte solution every 30 days b Remove the metal sensor guard 3 If installing a membrane for the first time remove the prote...

Страница 8: ... buffer value After 5 minutes and once the pH and temperature readings stabilize highlight Accept Calibration and press enter to confirm the second calibration point The message line will then display Ready for point 3 and you can continue with the 3rd calibration point if desired If you do not wish to perform a 3rd calibration point press Cal to complete the calibration V Ammonium sensor calibrat...

Страница 9: ...he instrument will use the internal barometer during calibration and will display this value in brackets at the top of the display Highlight Barometer and press enter to adjust it if needed If the barometer reading is incorrect it is recommended that you calibrate the barometer Note the barometer should be reading true barometric pressure see Barometer section of Instrument Manual for more informa...

Страница 10: ...low the sensor sufficient time to stabilize 4 pH sensor The key to pH sensor storage short or long term is to make certain that the sensor does not dry out Sensors which have been allowed to dry out due to improper storage procedures may be irreparably damaged by the dehydration and will require replacement You can try to rehydrate the sensor by soaking it preferably overnight in a potassium chlor...

Страница 11: ...Jessica M Barber Digitally signed by JESSICA BARBER Date 2020 03 10 11 22 25 04 00 ...