WILDLIFE 1001 Скачать руководство пользователя страница 3

IOP 019A.1 

Effective Date: 1/11/2017 

Page 3 of 11 

 

 

 

 

3. 

Sensors – Four sensors are required to measure all of the water chemistry 

parameters indicated above.  Sensors are installed in the quatro cable bulkhead 

ports indicated below to ensure proper function.  If measurement of one or more 

parameters is not required, the corresponding sensor(s) may be removed and 

replaced with a sensor port plug.  Accurate calculation of ammonia from 

measured ammonium requires installed and properly calibrated pH and 

temperature sensors. 

a. 

temperature/conductivity sensor (5560) installed in “TC” port 

b. 

pH sensor (1001) installed in ISE port 1 

c. 

ammonium sensor (1004) installed in ISE port 2 

d. 

polarographic dissolved oxygen sensor (2003)  installed in “DO” port 

II. 

Pro Plus Key Pad Functions 

Key 

Description 

 

System - 

Opens System Menu from any screen. Use to adjust system settings. 

 

Sensor - 

Opens Sensor Menu from any screen. Use to enable sensors and display 

units. 

Cal 

Calibrate - 

Opens Calibrate Menu from any screen. Use to calibrate all parameters 

except temperature. 

 

File - 

Opens File Menu from any screen. Use to view data and GLP files, set up site 

and folder lists, and delete data. 

 

Backlight - 

Press to turn the instrument backlight on and off and to adjust the 

display contrast when pressed with the left or right arrow key. 

 

Right Arrow - 

Use to navigate right in alpha/numeric entry screens. Can be pressed 

simultaneously with Backlight key to increase display contrast. 

 

Down Arrow - 

Use to navigate through menus and to navigate down in 

alpha/numeric entry screens. 

 

Power - 

Press to turn the instrument on or off. 

Help - 

Press to receive hints & tips during operation. 

ENTER 

 

Enter - 

Press to confirm selections, including alpha/numeric key selections. 

 

Left Arrow - 

Use to navigate left in alpha/numeric entry screens.  Press to return to 

previous menu in all screens except alpha/numeric entry. Can be pressed 

simultaneously with Backlight key to decrease display contrast. 

Esc 

Exit/Escape - 

Exits back to Run Screen. When in alpha/numeric entry screen, 

escapes to previous menu. 

 

Up Arrow - 

Use to navigate through menus and to navigate up in alpha/numeric 

entry screens. 

 

III. 

Meter preparation and maintenance 

Содержание 1001

Страница 1: ...ntrol Centre 1219 Queen Street East Sault Ste Marie Ontario P6A 2E5 Canada INSTRUMENT OPERATING PROCEDURE INSTRUMENT Multiparameter Water Quality Instrument MODEL Pro Plus Meter 6050000 Quatro Cable 605790 Probes temperature conductivity 5560 pH 1001 or 1001A DO 2003 ammonium 1004 MANUFACTURER Yellow Springs Instrument Co Inc YSI PRECAUTIONS POTENTIAL INTERFERENCES General Greatest accuracy is ach...

Страница 2: ...ulated from ammonium sensor measurements Exposure to the high ionic content of pH buffers can cause a significant but temporary drift in the ammonium sensors Therefore when calibrating the pH sensor remove ammonium sensor from the cable bulkhead and plug the port After pH calibration is complete replace the ISE sensors and proceed with ammonium calibration Alternatively the ammonium sensor may be ...

Страница 3: ... screen Use to calibrate all parameters except temperature File Opens File Menu from any screen Use to view data and GLP files set up site and folder lists and delete data Backlight Press to turn the instrument backlight on and off and to adjust the display contrast when pressed with the left or right arrow key Right Arrow Use to navigate right in alpha numeric entry screens Can be pressed simulta...

Страница 4: ... B Instrument software setup 1 Press the Power key to turn the instrument on The instrument will briefly display the splash screen with the YSI logo then go directly to the main run screen The first time the instrument is powered up or if the instrument has had a battery change with batteries removed for more than 2 minutes you will need to set the date and time a Highlight Date Time from the Syst...

Страница 5: ...ress enter 2 Highlight pH and press enter to enable it 3 Highlight pH mV and press enter to enable it 4 Press to return to the previous menu c Ammonium sensor 1 Highlight ISE2 NH4 Press enter 2 Highlight NH4 N mg l and press enter to enable it 3 Highlight NH4 mV and press enter to disable it 4 Highlight NH3 N mg l and press enter to enable it 5 Press to return to the previous menu d Dissolved oxyg...

Страница 6: ...etely without feeling any resistance Damage to your cable sensor may occur if you force the parts together e Once completely installed the nut will seat flat against the bulkhead At this point use the included tool to turn the nut an additional to turn DO NOT overtighten Refer to the sensor installation sheet that is included with the conductivity temperature sensor for detailed instructions 4 Ins...

Страница 7: ... clockwise to engage threads and finger tighten Do not use a tool This connection is waterproof Refer to the sensor installation sheet included with each sensor for detailed instructions 7 Install the 2003 DO sensor membrane as follows a Change membrane and oxygen probe electrolyte solution every 30 days b Remove the metal sensor guard 3 If installing a membrane for the first time remove the prote...

Страница 8: ... buffer value After 5 minutes and once the pH and temperature readings stabilize highlight Accept Calibration and press enter to confirm the second calibration point The message line will then display Ready for point 3 and you can continue with the 3rd calibration point if desired If you do not wish to perform a 3rd calibration point press Cal to complete the calibration V Ammonium sensor calibrat...

Страница 9: ...he instrument will use the internal barometer during calibration and will display this value in brackets at the top of the display Highlight Barometer and press enter to adjust it if needed If the barometer reading is incorrect it is recommended that you calibrate the barometer Note the barometer should be reading true barometric pressure see Barometer section of Instrument Manual for more informa...

Страница 10: ...low the sensor sufficient time to stabilize 4 pH sensor The key to pH sensor storage short or long term is to make certain that the sensor does not dry out Sensors which have been allowed to dry out due to improper storage procedures may be irreparably damaged by the dehydration and will require replacement You can try to rehydrate the sensor by soaking it preferably overnight in a potassium chlor...

Страница 11: ...Jessica M Barber Digitally signed by JESSICA BARBER Date 2020 03 10 11 22 25 04 00 ...

Отзывы: