6 Commissioning, operation
WIKA Operating Instruction, Model CPH7600 33
14051628.
01
07/
2012
G
B/
D
To use the "%-ERROR" function, perform the following:
1. Once the calibrator is switched on and operating, press the
F3
key to activate the "
MORE
"
menu option. Then press the
F1
key to activate the "
%-ERROR
" option.
2. Press the F1 key to select the "
CONFIG
" option.
3. The first option is the port setting. Use the "
SELECT
" option to scroll through the choice of
ports (pressure connections). Once this is completed, select the "
NEXT
" option.
4. "
LOOP POWER
" can be toggled on or off; select "
NEXT
" when done.
5. Use "
SELECT
" to scroll through the "
UNIT
" options, and select "
NEXT
" to move on.
GB