8 Faults / 9 Dismounting, return and disposal
WIKA Operating Instruction, Model CPH7600 37
14051628.
01
07/
2012
G
B/
D
8 Faults
Display
Reason
Measures
Low battery power, device will only
work for a short while longer.
Charge rechargeable battery
OL
-OL
Well above or below the limits of
the measuring range
Check: is the pressure within the permissible
measuring range of the sensor?
No display or
strange symbols,
device is not
responding to input
Battery flat
Charge rechargeable battery
System error
Switch off the CPH7600, wait for a short
period of time, switch on again
Technical fault
Return to manufacturer for repair
CAUTION!
If faults cannot be eliminated by means of the measures listed above, the hand-held
pressure calibrator must be shut down immediately, and it must be ensured that
pressure and/or signal are no longer present, and it must be prevented from being
inadvertently put back into service. In this case, contact the manufacturer. If a return is
needed, please follow the instructions given in chapter "9.2 Return".
9 Dismounting, return and disposal
9.1 Dismounting
Only disconnect test- and calibration installations once the system has been depressurised!
9.2 Return
WARNING!
Strictly observe when shipping the instrument:
All instruments delivered to WIKA must be free from any kind of hazardous
substances (acids, bases, solutions, etc.).
When returning the instrument, use the original packaging or a suitable transport
package.
To avoid damage:
1. Wrap the instrument in an antistatic plastic film.
2. Place the instrument, along with the shock-absorbent material, in the packaging. Place shock-
absorbent material evenly on all sides of the shipping box.
3. If possible, place a bag, containing a desiccant, inside the packaging.
4. Label the shipment as transport of a highly sensitive measuring instrument.
GB
Содержание Wally Box III
Страница 10: ...3 Specifications 10 WIKA Operating Instruction Model CPH7600 14051628 01 07 2012 GB D 3 Specifications GB ...
Страница 11: ...3 Specifications WIKA Operating Instruction Model CPH7600 11 14051628 01 07 2012 GB D GB ...
Страница 17: ...4 Design and Functions WIKA Operating Instruction Model CPH7600 17 14051628 01 07 2012 GB D GB ...
Страница 47: ...3 Technische Daten WIKA Betriebsanleitung Typ CPH7600 47 14051628 01 07 2012 GB D Sensor und Kabel D ...
Страница 53: ...4 Aufbau und Funktion WIKA Betriebsanleitung Typ CPH7600 53 14051628 01 07 2012 GB D D ...