8
WIKA operating instructions, model MG-1
11609681.03 08/2019 EN/DE/FR/ES
EN
3. Specifications
3. Specifications
3.1 Measuring ranges
Gauge pressure
bar
Measuring range
0 ... 6
0 ... 10
0 ... 16
0 ... 200
0 ... 300
0 ... 400
Overload safety
20
20
32
500
800
800
Burst pressure
25
25
160
1,200
1,700
1,700
psi
Measuring range
0 ... 100
0 ... 150
0 ... 200
0 ... 3,000
0 ... 4,000
0 ... 5,000
Overload safety
290
290
460
7,200
11,000
11,000
Burst pressure
1,450
1,450
2,300
17,000
24,000
24,000
Vacuum tightness: Yes
3.2 Output signals
Signal type
Signal
Current (2-wire)
4 ... 20 mA
Voltage (3-wire)
DC 0 ... 10 V
DC 0 ... 5 V
DC 1 ... 5 V
Ratiometric (3-wire)
DC 0.5 ... 4.5 V
Depending on the signal the following loads apply:
Signal
Load in Ω
4 ... 20 mA
≤ (power supply - 8 V) / 0.02 A
DC 0 ... 10 V
> max. signal / 1 mA
DC 0 ... 5 V
> max. signal / 1 mA
DC 1 ... 5 V
> max. signal / 1 mA
DC 0.5 ... 4.5 V ratiometric
> max. signal / 1 mA