WIKA operating instructions, models CTH6300 and CTH6500
EN
27
2079988.08 02/2018 EN/DE/IT
3.
Change the 4 characters using the ▲ arrow key.
4.
Press the ▼ arrow key to go to the next position.
5. When all 4 characters are entered as required, confirm by pressing
ENTER/MENU
.
⇒
A small
2
appears and the following 4 characters can be changed as well.
6. Use
ESC
to return to the measuring mode.
When accessing
oP1
using
ENTER/MENU
, the
oP1
function (calibration by
code) is activated, even if the menu has been exited by pressing
ESC
.
Example of the measured value display after entering a probe calibration code:
The small
1
in the left corner, in combination with the display segment
CAL
in the centre
of the display, indicates that the calibration by code
oP1
has been activated.
Physical calibration oP2
1.
Use the ▲▼ arrow keys to select
oP2
.
2. Press
ENTER/MENU
to confirm the desired setting.
⇒
1 P
is displayed in the lower part of the display.
3.
Use the ▲▼ arrow keys to select from 1-point
1 P-
, 2-point
2 P-
and 3-point
3
P-
calibration.
oP2
1 P
2 P
3 P