9037926
07/2015
60
WIKA Operating Instructions
CTD9300
Rev. 1
Temperature Dry Well Calibrators
CTD9300
5. Technical Data
GB
5. Technical Data
5.1
CTD9300
from -30 to 165 °C
Minimum temperature
1)
Maximum temperature
Error of measurement up to 100 °C
Error of measurement above 100 °C
Stability up to -30 °C
Stability at 165 °C
Gradients axially, 4 cm, 165 °C
Gradients axially, 4 cm, -30 °C
Heating time from 20 °C to 165°C
Cooling time from 20 °C to -20 °C
Immersion depth
Sleeve diameter
Resolution in K
Footprint W x D
Footprint
Dimensions, W x D x H
Power consumption
Power supply
Fuse
Measurement value logging
Mass (without integ. measuring
instrument)
Ambient temperature
Standard ambient conditions
Overvoltages
Degree of pollution
Storage temperature
Humidity
Protection class
CE conformity
°C
°C
K
%
K
K
K
K
min
min
mm
mm
K
mm
cm²
mm
W
VAC
kg
°C
°C
% rel.h.
-30
165
0.1
0.1 from the measured value
< 0.05
0.01
< 0.06
< 0.04
12
7
160
28
0.01
160 x 320
512
160 x 320 x 420
400
100 ... 240 +10 %/-15 %
50/60 Hz
6.3A, slow blowing (T6.3A 250V)
optional
10
0...40
Use permissible indoors only
permissible altitude up to
2,000 m
Overvoltage category II
Transient overvoltage as it
normally occurs in the supply
network
Degree of pollution 2
-20 ... 40
0 ... 90
IP 20
EN61326-1, EN61010-1
1 )
1 )
1 )
1 )
1 )
at 23 °C ambient temperature