Operating Instructions CPG2300 Series
35
Portable Digital Pressure Gauge
CPG2300 Series
Cal Zero and Cal Span Screens: Digits are selected by using the right or left arrow
keys. The selected digit has a cursor below it. Adjustments of the digit is by the up or
down arrow keys. Current input zero and span adjustments are limited to 5% of the
range (16 mA). This would be +/- 0.8 mA. Press enter key to store the change.
mA Input
Zero Calibration
+4.00
1
mA
Enter True Current
8.1.2 Operational Notes
When the current measurement option is enabled the value is displayed as the alternate
channel to the pressure channel that is selected. This means when “Sensor A” is se-
lected as the pressure measurement channel the pressure will be displayed as the top
reading and the current measurement will be the bottom reading.
The display order is reversed when “Sensor B” is selected.
Both pressure channels can also be displayed while the current measurement channel
is active by selecting “Sensor A&B”. The current measurement can then be acquired
over the serial port. This is accomplished by issuing the “I?” command (see Remote
Operation section).
The peak pressure indication function can not be enabled when the
current input is enabled. The null feature will work when the current
input is enabled.
mA Input
Zero Calibration
+20.002
mA
Enter True Current
Содержание CPG2300 Series
Страница 8: ...8 Operating Instructions CPG2300 Series Portable Digital Pressure Gauge CPG2300 Series Notes...
Страница 32: ...32 Operating Instructions CPG2300 Series Portable Digital Pressure Gauge CPG2300 Series Notes...
Страница 38: ...38 Operating Instructions CPG2300 Series Portable Digital Pressure Gauge CPG2300 Series Notes...
Страница 43: ...Operating Instructions CPG2300 Series 43 Portable Digital Pressure Gauge CPG2300 Series Notes...