WIKA operating instructions EMM Calibration Sled 

PN 0019408001A  08/2018  EN

 EN

1. Setup

2.  Program operation and procedures

2.1  Calibration Procedures

1. Connect EMM to sled as shown on page 3. Apply power. 

2. Start EMM Calibration program. It should appear as shown in Figure 2.1. 

Figure 2.1 – Main screen

3. Set Com Port to your computers com port value. The baud rate settings are built into the application. They are 57600, N, 8, 1. 

The Com Test button brings up the window below. If the EMM is not connected or a problem exists an error will be shown. 

 

 

    

                    

 

   

   

   Figure 2.2 A – Successful com test  

 

 Figure 2.2 B - Communication error

 

1.2  Software installation

To install the software, run the setup program on the included disk (setup.exe). The setup program will install all the necessary 

files to run the program. It will register all files appropriately so it can be completely removed using the Windows® add/remove 

program. Upon installation, a desktop shortcut and start menu shortcut will be added.  If this is not necessary, the “Xducer_Utility.

exe” program may be run directly. 
The computer on which the software is run must have Windows® XP or greater. It must have an available RS-232 serial port 

capable of running at 57600 baud.  Finally, Microsoft .NET Framework 4.0 or greater must be installed. This is included as the file 

“dotNetFx40_Full_x86_x64.exe”. The latest version can be downloaded from: http://www.microsoft.com/net
Calibration Sled Hardware Setup: Connect the AC adapter of the power supply to an AC power socket. Then connect the power 

supply to the circular power connector on the rear of the calibration sled. Connect the 9 pin d-sub connector on the calibration sled 

to any available RS-232 connector on the computer. This connection is made through the straight 9 pin male to 9 pin male cable.
Transducer Com Cable Setup: Connect the AC adapter of the power supply to an AC power socket. Then connect the “Computer” 

side of the cable to any available RS-232 connector on the computer. Connect the “Transducer” side of the cable to the EMM 

communication port.

Содержание CPD8500

Страница 1: ...s reserved Mensor is a registered trademark of Mensor LP All other brand and product names are trademarks or registered trademarks of their respective companies Prior to starting any work read the operating instructions Keep for later use Transducer utility manual 0020646001 ...

Страница 2: ...A 08 2018 EN EN Contents Contents 1 Introduction 3 1 1 Equipment required 3 1 2 Technincal information 3 2 Program operation and procedures 4 2 1 Calibration Procedures 4 2 2 Software installation 4 2 3 Use of multiple calibration methods 7 2 4 Additional features 8 ...

Страница 3: ...n platinum resistive thermometer PRT The rated accuracies are Initial Accuracy Stability Ambient Temperature 4 C 2 C year Relative Humidity 3 2 year Barometric Pressure 25 2 year PRT 2 C 1 C year It is important to use standards that have a test uncertainity ratio of at least 1 1 A higher ratio is better 1 3 Equipment required EMM Calibration Sled P N 0019780001 EMM P N 0019582001 CPD8500 EMM Cal ...

Страница 4: ...a desktop shortcut and start menu shortcut will be added If this is not necessary the Xducer_Utility exe program may be run directly The computer on which the software is run must have Windows XP or greater It must have an available RS 232 serial port capable of running at 57600 baud Finally Microsoft NET Framework 4 0 or greater must be installed This is included as the file dotNetFx40_Full_x86_x...

Страница 5: ...head temperature measurement that has an external probe Ambient Temperature is the internal ambient temperature sensor Ambient Pressure is the internal barometric pressure sensor Relative Humidity is the internal humidity sensor Observe each reading to determine if sensors are live 6 The PRT should be calibrated by probe immersion in a precision temperature bath against a standard that has accurac...

Страница 6: ... entering the reference value then selecting the Take Reading button A graph is updated as the points are entered When the table is fully populated select Calibration and Save Calibration The linearity correction will be applied and saved Figure 2 5 Linearity correction graph Other options are available such as reducing the number of calibration points and upscale downscale calibrations By changin...

Страница 7: ...e used to change the slope This value is nominally 1 This is a multiplier such that 1 would not change the current slope A value of 0 9 would multiply the current slope by 0 9 or 90 The Zero and Span window is shown in Figure 4 Figure 2 6 Zero and Span window 2 pt cal Use of the 2 pt Cal button requires the capability to apply two discrete calibration stimuli to the individual sensors A reference ...

Страница 8: ...A Alexander Wiegand SE Co KG Alexander Wiegand Straße 30 63911 Klingenberg Germany Tel 49 9372 132 0 Fax 49 9372 132 406 info wika de www wika de Mensor Corporation 201 Barnes Drive San Marcos TX 78666 USA Tel 1 512 3964200 15 Fax 1 512 3961820 E Mail sales mensor com www mensor com ...

Отзывы: