background image

 

Technical Specifications (Continuation) 

 

Piston-Cylinder Assemblies

 

Standard Masses 

  Material: tungsten carbide 

  Poisson’s ratio: 0,218 

  Young’s modulus: 6 10

11 

N/m 

Typical geometry researched: 

  Straightness (typical manufacturing tolerance): 0,1 µm 

  Roundness (typical manufacturing tolerance): 0,1 µm 

  Parallelism (typical manufacturing tolerance) : 0,1 µm 

  Clearance between the piston and the cylinder: 0,2 to 0,4 

µm according to the model 

  Stability of the effective area: 

 1 ppm / an  

Internal mass 

  Material: nickel chromium steel  

  Mass density: 7900 kg/m

3

 ± 10% 

 

External optional calibration masses 

  Material: 304L steel 

  Mass density: 7920 kg/m

3

± 10%  

  Composition of the set: 5 x 2 kg 

  Adjusting tolerance to the nominal  value: ±1E

-6  

  Calibration Uncertainty: ± 2,4E

-6  

EMM Environmental Measuring Module

 

Sensors 

Type 

Accuracy 

Alarm setting 

Ambient temperature: 

4-wire PT100 

±

 0,2°C 

±

 2°C 

Relative humidity: 

Capacitive sensor 

±

 5 % 

±

 20% 

Atmospheric pressure: 

Strain gauge 

±

 2 mbar 

±

 10 mbar 

PCA temperature: 

A DIN 43760 4-wire PT100  

± 0,1 °C 

N/A 

Metrological Specifications

 

Researched metrological specifications 

Linearity: 2E

-6

 FS 

Hysteresis: 2E

-6

 FS 

Bias: 7E

-6

 FS 

Repeatability: 

 5E

-6

 FS 

Temperature effect: full compensation 

Precision* : 

 11E

-6

 FS (

Combination of repeatability, hysteresis, linearity and 3-year stability) 

Typical uncertainty researched for a 200 kPa range (k=2) : U=0.8 Pa + 1.E-5 P 

 

Dimension (L x W x H): 

530 x 400 x 320  mm 

Measurement fluid: 

Dry and non corrosive clean gas 

Maximum overpressure: 

110 % FS 

Operating temperature: 

10 to 30 ° C 

Sampling rate: 

250 ms 

Operating humidity: 

15 to 85 % HR 

Computer interface: 

RS232C 

Power supply: 

110 - 240 VAC, 50-60 Hz 

Residual vacuum : 

Pirani gauge 

 1Pa ±1E

-4

 

Power consumption: 

20 VA 

Weight 

20 kg 

 

 

 

 

 

 

Summary of the Functions

 

The CPD8000 has a great variety of functions aimed at facilitating its use and assuring the quality of its long-term 
operation: 

Menus in English, French, German and Spanish 

Saving of the metrological characteristics for 3 pressure 
ranges 

Large LCD back-lighting graphic screen with contrast adjustment

 

Mechanical protection against overpressure up to 110% 

Overpressure visual and sound warning

 

Calibration visual warning when the calibration is 
necessary 

Pressure display in 12 pressure units, and a user unit 

User adjustable pressure stability criterion 

Front panel tarring key 

Front panel ACF running key 

Choice between internal/external calibration 

EMM parameters display 

Piston-cylinder assembly temperature measuring circuit 
with auto-calibration by resistance of integrated reference 

Permanent display of the residual vacuum 

 

 

Summary of the Functions (Continuation) 

 

Page 7 of 8 

 

 

 

 

                  Desgranges & Huot WIKA Data Sheet CPD8000 04/2012 

Содержание cpd8000-g

Страница 1: ...ea CPD8000 is a worldwide exclusivity combining two high level technologies Best available piston cylinder assemblies whose effective area A is perfectly known High accuracy force cell which measure t...

Страница 2: ...only Dedicated to gauge measurements the force cell is let at the atmospheric pressure CPD8000 A Absolute and gauge mode Versatile measurer the force cell is placed in reference chamber which can be a...

Страница 3: ...the force transmitted to the measuring force cell by the pressure is perfectly and fully vertical then without alteration It has been created to make sure that the rotating noises be less important th...

Страница 4: ...se masses can also be used to run an external calibration 5 The Environment Monitoring Module EMM In order to determine whether it is worthy to use the ACF the CPD8000 is equipped with an Environment...

Страница 5: ...olatile memory of the CPD8000 Kn of the piston cylinder assembly Sensitivity of the measuring force cell Nk Normal gravity gn Local gravity gl Pressure distortion coefficient of the piston cylinder as...

Страница 6: ...kg Pure gas A06 5 Pa 5 MPa 5 Pa 5 bar kg Pure gas G01 0 1 Pa 100 kPa 0 1 Pa 0 1 bar kg Pure gas G02 0 2 Pa 200 kPa 0 2 Pa 0 2 bar kg Pure gas G03 0 5 Pa 500 kPa 0 5 Pa 0 5 bar kg Pure gas G04 1 Pa 1...

Страница 7: ...ity hysteresis linearity and 3 year stability Typical uncertainty researched for a 200 kPa range k 2 U 0 8 Pa 1 E 5 P Dimension L x W x H 530 x 400 x 320 mm Measurement fluid Dry and non corrosive cle...

Страница 8: ...rmances and the calculation of its measurement uncertainty by comparison with a reference standard The CPD8000 complies with the following European directives and norms n 89 336 CEE Electromagnetic Co...

Отзывы: