WIKA operating instructions, model CEP1000
11570351.02 10/2010 GB/D/F/E/I
GB
6. Commissioning, operation
6.3.3 Data acquisition for minimum and maximum
If the CEP1000 is operated in a measuring mode (milliampere, milliampere with power
supply or voltage), the minimum and maximum values for the corresponding input mode
are monitored and stored permanently. To activate the data acquisition proceed as
follows:
1) Switch the CEP1000 to a measuring mode.
2) Connect the instrument to the instrument to be tested.
3) Press both keys (“MIN” and “MAX”) to clear the memory for the minimum and
maximum values. The “CLR” message is displayed on the LCD display, if the
memory has been deleted.
4) Acquire the minimum value and the maximum value using the instrument as long as
required.
If necessary, deactivate the battery saving function prior to calling the
data acquisition function to avoid premature switching off of the calibra
-
tor.
5) The saved value can be accessed any time by pressing the "
MIN
" or "
MAX
" key.
6) When switching off the instrument or switching to another operating mode the
memory will be cleared. Back up the data before switching off.
Содержание CEP1000
Страница 24: ...24 WIKA operating instructions model CEP1000 11570351 02 10 2010 GB D F E I GB...
Страница 46: ...46 WIKA Betriebsanleitung Typ CEP1000 D 11570351 02 10 2010 GB D F E I...
Страница 68: ...68 WIKA mode d emploi type CEP1000 F 11570351 02 10 2010 GB D F E I...
Страница 90: ...90 WIKA manual de instrucciones modelo CEP1000 E 11570351 02 10 2010 GB D F E I...