![WIKA APGT43.100 Скачать руководство пользователя страница 31](http://html1.mh-extra.com/html/wika/apgt43-100/apgt43-100_operating-instructions-manual_980121031.webp)
31
Pressure gauges with output signal, optionally for hazardous areas
11297264.05 07/2019 EN/DE
EN
9. Specifications
Models DPGT43.100, DPGT43.160
Pressure limitation
Steady
Full scale value
Fluctuating
0.9 x full scale value
Overload safety and
max. working pressure
see the following table
Temperature effect
When the temperature of the measuring system deviates from
the reference temperature (+20 °C): max. ±0.5 %/10 K of full
scale value
Wetted materials
Media chamber with process
connection, venting of the media
chambers, bellows
Stainless steel 316Ti (1.4571)
Pressure element
≤ 0.25 bar: Stainless steel 316L
> 0.25 bar: NiCr alloy (Inconel)
Non-wetted materials
Case, bayonet ring
Stainless steel
Movement
Brass
Dial
Aluminium, white, black lettering
Pointer
Aluminium, black
Window
Laminated safety glass
Ingress protection per IEC/EN 60529
IP54; option: IP65
Overload safety and max. working pressure
Scale ranges
Overload safety in bar
Either side max.
max. working pressure in bar
(static pressure)
Standard
Option
Standard
Option
0 ... 16 to 0 ... 40 mbar
2.5
-
2.5
6
0 ... 60 to 0 ... 250 mbar
2.5
6
6
10
0 ... 400 mbar
4
40
25
40
0 ... 0.6 bar
6
40
25
40
0 ... 1 bar
10
40
25
40
0 ... 1.6 bar
16
40
25
40
0 ... 2.5 to 0 ... 25 bar
25
40
25
40
For further specifications see data sheet PV 17.05