![WIKA APGT43.100 Скачать руководство пользователя страница 26](http://html1.mh-extra.com/html/wika/apgt43-100/apgt43-100_operating-instructions-manual_980121026.webp)
Pressure gauges with output signal, optionally for hazardous areas
26
11297264.05 07/2019 EN/DE
EN
9. Specifications
DANGER!
Danger to life due to loss of explosion protection
The non-observance of the instructions for use in hazardous areas can lead to
the loss of the explosion protection.
▶
Adhere to the following limit values and instructions.
▶
The case contains parts made of non-metallic material, which could be
adversely affected by environmental conditions. It must be ensured that the
non-metallic parts are not damaged as a result of the environmental condi-
tions.
Non-metallic parts of the case
PUR, TPU, PA 6, EPDM, NBR (option), FMQ (option)
9.1 Electrical connection
Standard cable socket
Cable socket, fitted on the right-hand side of the case.
Material: PA 6, black
Per VDE 0110 insulation group C/250 V
Cable gland M20 x 1.5 (facing downwards) with strain relief
6 screw ter PE for conductor cross-section 1.5 mm²
Cable outlet
Cable length 2 m or 5 m, facing downwards and to the right
Material: PVC
Miniature connector
Miniature connector M8 x 1, 4-pin