![WIKA A-RB-1 Скачать руководство пользователя страница 22](http://html.mh-extra.com/html/wika/a-rb-1/a-rb-1_operating-instructions-manual_980436022.webp)
22
4.2 Programming of parameters
Com-
Response
Meaning
mand
No. of byte
Example
PMODE
1 = REAL mode
signal to set
2 = HOLD mode
mode of instrument
3 = MIN mode
4 = MAX mode
5 + 1 byte +
↵
PMODE1
PMINM
0 = Reset
erase MIN memory
5 + 1 byte +
↵
PMINM0
PMAXM
0 = Reset
erase MAX memory
5 + 1 byte +
↵
PMAXM0
PST1E
SE of alarm 1
set switch-on value alarm 1, incl.
5 + 6 byte +
↵
sign and decimal point
PST1E+10.00
PST1A
SA of alarm 1
set switch-off value alarm 1, incl.
5 + 6 byte +
↵
sign and decimal point
PST1A+09.50
PST2E
SE of alarm 2
set switch-on value alarm 2, incl.
5 + 6 byte +
↵
sign and decimal point
PST2E+08.00
PST2A
SA of alarm 2
set switch-off value alarm 2, incl.
5 + 6 byte +
↵
sign and decimal point
PST2A+08.50
PSCLA
Low scale
set low end of scale value incl.
5 + 6 byte +
↵
sign and decimal point
PSCLA+00.00