background image

WIKA Alexander Wiegand SE & Co. KG

Alexander-Wiegand-Straße 30

63911 Klingenberg/Germany

Tel.  +49 9372 132-0

Fax  +49 9372 132-406

[email protected]

www.wika.de

2408976.06  12/2016

Operating 

instructions

1. Safety

WARNING!

Before installation, commissioning and 

operation, ensure that the appropriate 

pressure gauge has been selected in 

terms of measuring range, design and 

suitable wetted material (corrosion) for the specific 

measuring conditions. In order to guarantee the 

measurement accuracy and long-term stability 

specified, the corresponding load limits must be 

observed.

Only qualified persons authorised by the plant 

manager are permitted to install, maintain and 

service the pressure gauges.
For hazardous media such as oxygen, acetylene, 

flammable or toxic gases or liquids, and refrigera

-

tion plants, compressors, etc., in addition to all 

standard regulations, the appropriate existing 

codes or regulations must also be followed.
From pressure gauges which do not correspond 

to a safety version per EN 837 highly pressurised 

media might leak out through the possibly burst-

ing window in case of a component failure. For 

gaseous media and working pressures > 25 bar a 

pressure gauge with safety version S3 is recom-

mended per EN 837-2.

After an external fire, pressure media can leak out, 

particularly at soft solder joints. All instruments 

must be checked and, if necessary, replaced 

before recommissioning the plant.
Non-observance of the respective regulations 

can cause serious injuries and/or damage to the 

equipment.

2. Mechanical connection

In accordance with the general technical regula-

tions for pressure gauges (e.g. EN 837-2). When 

screwing the instruments in, the force required to 

do this must not be applied through the case, but 

only through the spanner flats provided for this 

purpose, and using a suitable tool.

For parallel threads, use flat gaskets, lens-type 

sealing rings or WIKA profile sealings at the 

Examples:

Model 213.40

Model 732.14, 

overpressure safety 

up to 400 bar

Model 432.56, 

overpressure safety 

up to 100 bar

Pressure gauges

Notes per current pressure equipment directive

 

„

The pressure gauges are defined as “pressure 

accessories”

 

„

The volume of the “pressure-bearing housings” of 

WIKA pressure gauges is < 0.1 L

 

„

The pressure gauges carry CE marking for fluid 

group 1 per annex II, diagram 1 when their permis-

sible working pressure is > 200 bar

Instruments that do not carry the mark are manufactured 

per article 4, paragraph 3 “sound engineering practice”.
Applicable standards (depending on model)

EN 837-1  Bourdon tube pressure gauges, dimensions, 

metrology, requirements and testing

EN 837-2  Selection and installation recommendations 

for pressure gauges

EN 837-3  Diaphragm and capsule pressure gauges, 

dimensions, metrology, requirements and 

testing

Specifications: See data sheet at www.wika.de
Subject to technical modifications.

© WIKA Alexander Wiegand SE & Co. KG 2009

EN

sealing face 

. With tapered threads (e.g. NPT 

threads), sealing is made in the threads 

 using 

additional sealing materials, e.g. PTFE tape 

(EN 837-2).

The torque depends on the sealing used. In order 

to orientate the measuring instrument so that it 

can be read as well as possible, a connection with 

clamp socket or union nut should be used.

When a blow-out device is fitted to a pressure 

gauge, it must be protected against being blocked 

by debris and dirt. With safety pressure gauges 

(see 

k

) there must be a free space of > 15 mm 

behind the blow-out back.

After installation, open the vent valve (if available) 

or set from CLOSE to OPEN. With 

models 4 and 7, do not open the 

flange mounting screws. The version 

of the vent valve depends on the 

model and can deviate from the above illustration!

Requirements for the installation point

If the line to the measuring instrument is not 

adequately stable, a measuring instrument holder 

should be used for fastening (and possibly via a 

flexible capillary). If vibrations cannot be avoided 

by means of suitable installation, instruments 

with liquid filling should be used. The instruments 

should be protected against coarse dirt and wide 

fluctuations in ambient temperature.

Note for model 732.14, for front bezel mounting: 

The front bezel serves as centring and as the 

aperture in the mounting panel. Securing and thus 

the weight-bearing must be made via the pressure 

connection piping.

3. Permissible ambient and operating 

temperatures

When mounting the pressure gauge it must be 

ensured that, taking into consideration the influ

-

ence of convection and heat radiation, no deviation 

above or below the permissible temperature limits 

can occur. Observe the influence of temperature 

on the indication accuracy!

4. Storage

To protect the pressure gauges from mechanical 

damage keep them in the original packaging until 

installation.

Protect the measuring instruments from humidity 

and dust.

Storage temperature range: -40 ... +70 °C

Storage temperature range model 

PG23LT: -70 ... +70 °C

5. Maintenance and repairs

The pressure gauges are maintenance-free. 

Regular checks should be carried out to ensure 

the measurement accuracy. Checks or recalibra-

tions must only be carried out by qualified skilled 

personnel with the appropriate equipment. When 

dismounting, close the vent valve (if available).

WARNING! 

Residual media in 

dismounted pressure gauges can 

result in a risk to persons, the environ-

ment and equipment. Take sufficient 

precautionary measures. 

Sealing in 

the thread 

Spanner flats

Sealing face 

Installation with 

open-ended 

spanner
Sealing of the pressure gauge connections

Отзывы: