background image

Inteo
Test
Mode

The Test Mode allows the clinician to compare the output of the hearing aid against factory
specifications. The hearing aid is set to a fixed frequency gain setting during this measurement.

1. Turn  off  the  iP5  by  pressing  [

ON  R

]  AND

[

ON L

] simultaneously.

2. Connect the Inteo hearing aid to the appro-

priate inlet [

ON L or ON R

].

3. Hold  down  the  top  right  menu  key  while

pressing the appropriate [

ON

] key.

4. [

TEST MODE

] is displayed. The hearing aid

is in test mode.

5. Connect  Inteo  to  the  coupler  of  the  hearing  aid  test  system  and  follow  ANSI-1996  for 

evaluating the Inteo hearing aid.

6. Press [

TELE

] to evaluate telecoil function.

7. When testing a BTE instrument open the battery door, when testing is completed, to restore

the hearing aid to regular functioning.

Programming
the
RC3
Remote
Control

1. Insert the batteries in the remote control and replace the compartment cover.

2 . Within 10 seconds, press and hold down the center key while pressing the upper left key (+).

3. The word Pro will appear in the display window.

4. Press the center key. The display window will show the RC3 model.

5. Press the center key. The display window will show the maximum number of programs the

remote can handle. Change the number of programs by using the up/down keys on the right
side of the remote. The number of programs in the remote must be the same number of pro-
grams in the hearing aid.

6. Press the center key. The display window will show the setting of the short-term log func-

tion. The default setting is off. Change the setting using the up/down keys on the right side
of the remote. Turning it on will enable the short-term logging feature of the aid. 

7. To enable separate adjustment of the right and left hearing aids (RC3-2 only), press the cen-

ter  key.  The  display  window  will  show  the  current setting. The default setting is "Uni,"
which means the remote will adjust both aids simultaneously Use the up/down arrow keys
on the right side to change the setting to "Sep," which will adjust each aid individually.

8. To enable the timer function (RC3-2 only), press the center key. The display will show the

current setting of the timer function. The default setting is "AL OFF." Change the setting
using the up/down arrow keys on the right side of the remote.

9. To  set  the  RC3 options,  press  the  center  key.  The  display  window  will  show  "Opt."  All

options are set to default values, and it is normally not necessary to change these during
the initial programming of the RC3.

10. To end programming, press the center key again.

B

12

IN_IP5_Manual8.qxd  11/28/06  1:12 PM  Page 12

Содержание iP5

Страница 1: ...1 800 221 0188 www widexPro com Fine Tuning Guide for Programming the Inteo with the iP5 Portable Programmer DIB12 1106 ...

Страница 2: ...s 4 Turn on iP5 Press on R for right or on L for left See A Note 1 Only one ear can be fit at a time In addition adjustments cannot be made nor data entered when the hourglass is active Note 2 An Inteo hearing aid that has been programmed and fine tuned with the iP5 program mer is completely compatible with Compass An Inteo hearing aid that has been programmed and fine tuned with Compass may not b...

Страница 3: ...Minimum Gain MIN values representing the minimum acceptable gain for a conversational input Press BACK see B to return to the Feedback window Inteo programming is complete Note Once the Inteo hearing aid has been programmed the FINE TUNE window will be the next to appear when the mode button is pressed FINE TUNE will be the first window to appear when the iP5 programmer is turned on and a programm...

Страница 4: ...ents distortion of high outputs It is recom mended that AOC remain active unless the client requires higher output from the hear ing aid even at its maximum gain settings Automatic Output Control may be turned off manually Fine Tuning Inteo 1 Turn on iP5 Press on R for right or on L for left See A Note Only one ear can be fine tuned at a time 2 Press MODE if iP5 is not in FINE TUNE mode See B When...

Страница 5: ...edure to add or change a listening program selection in each available program position To access Automatic Output Control press in the FINE TUNE mode followed by MPO Press AOC see B Press UP ARROW beneath OFF Press BACK see B to return to MPO An exclamation point will appear in the mode window next to AOC to indicate that a change has been made to Automatic Output Control Note When Automatic Outp...

Страница 6: ...ting e Press BACK see B to return to PRO GRAM OPTIONS An exclamation point will appear in the mode window next to PRG to indicate the selection of additional listening programs 2 Remote Control RC3 matching can be accomplished in the PROGRAM SELECTION mode a Program the RC3 remote control see Programming the RC3 Remote Control instructions b Press RC 3 in the PROGRAM SELEC TION mode window see B c...

Страница 7: ...e Left Push the ON button of the hearing aid you are program ming i e ON L to deactivate the binaural mode 5 The Speech and Noise mode settings can be adjusted for each acoustic listening program except Audibility Extender in the SPEECH NOISE mode Note Audibility Extender is displayed as AE in this screen to indicate that the listening program has been selected but no change to the speech and nois...

Страница 8: ...um number of programs the remote can handle Change the number of programs by using the up down keys on the right side of the remote The number of programs in the remote must be the same number of pro grams in the hearing aid 6 Press the center key The display window will show the setting of the short term log func tion The default setting is off Change the setting using the up down keys on the rig...

Отзывы: